Appendix A: Product Specifications
19
Appendix A: Product Specifications
Accuracy (non-linearity, hysteresis, and non-
repeatability)
± 0.5% of Reading (optional ± 0.15%)
± temperature coefficients
CE Mark Compliance
1
EMC Directive 89/336/EEC
Fittings
Standard
Optional
½” (12.7 mm) tubulation;
Cajon
®
8-VCR
®
only on 5K and higher ranges
KF-16, Cajon 8-VCR, Cajon 8-VCO
®
Full Scale Ranges
10, 100, 1K, 5K, 10K, 15K, 20K,
25K mmHg
Input Power Required
±15 VDC @ 35 mA, regulated ±2%
Materials Exposed to Gases
Inconel and stainless steel
Operating Temperature Range
0° to 50° C (32° to 122° F)
Output Signal
0 to 10 VDC into 10K ohm or greater
Overpressure Limit
120% of F.S. or 35 psia (240 kPa),
whichever is greater
Temperature Coefficients
Zero
Span
0.008 % F.S./° C
0.04% Reading/° C
Volume (Px side)
7.00 cc for 10, 100, 1K mmHg
8.5 cc for all other ranges
Due to continuing research and development activities, these product specifications are subject to change without notice.
1When used with optional metal, braided, shielded cables.
Summary of Contents for Baratron 122A Series
Page 4: ...Table of Contents iv Appendix A Product Specifications 19 Appendix B Product Code 21 Index 23...
Page 6: ...List of Figures List of Tables vi...
Page 12: ......
Page 13: ......
Page 16: ...Setup Chapter Two Installation 10 This page intentionally left blank...
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ...Signal Conditioner Electronics Chapter Three Overview 14 This page intentionally left blank...
Page 24: ...Sensor Chapter Five Maintenance 18 This page intentionally left blank...
Page 26: ...Appendix A Product Specifications 20 This page intentionally left blank...