946 Controller Instruction Manual
pg. 51
5.0E+2
Pressure Control Parameter Setting Recipe
Prop - Kp
P Sp
1.0E+1
Integral - Ti
Derivative - Td
1.0E+1
1.5E+0
90.0%
Ceiling
0.0%
Start
End
Preset
30%
Direction
Upstream
R1
R2
R3
R4
R5
R6
R7
R8
B1
B2
C1
C2
Rat
VLV
NA
NA
GainSchedBand
0%
SchedGainCoeff
1
C1
PID Ch
P Ctrl Ch
A1
30.0%
CtrlStart
10s
Base
0.0%
Figure 6-14 Single MFC Control Recipe Setup Screen
The Base, Ceiling, Preset, Start and End are the percentage of the full-scale flow
rate of the MFC on the assigned PID channel, not the actual flow rate.
6.7.3.1.1 Set the Recipe for Single MFC PID Control
To edit and set a PID recipe, following steps are required to enter the PID recipe editing mode:
1. Press the
System
Setup
button, move the cursor to PID Recipe Edit (see Figure 6-4), then press
Enter
key to display the Pressure Control Parameters Setting Recipe screen shown in
2. Press
or
keys to select a desired recipe for editing or displaying (highlighted by
grey background, such as C1 in Figure 6-14), then, press the
Enter
key to begin the
setup or edit mode. Any of the14 PID parameters and settings associated with a recipe
may now be entered or changed.
Only 4 parameters (P set point, Prop-Kp, Integral-Ti, and Derivative-Td) can be
modified while a PID recipe is actively running. The PID control must be stopped to
change any other parameters in an active recipe.
For non-active recipes, all parameters can be changed (either from front panel or serial
communication).
See Figure 6-14 for setting up MFC PID control recipe:
•
PID Ch: PID Ch: This selects either the channel of the controlling MFC (single MFC
control) or method for automatically controlling the system pressure (ratio or valve).
a) When single MFC is used for pressure control, select and confirm with
Enter
the channel where the MFC is connected.
b) When multiple MFCs are used for Ratio pressure control, select Rat and
confirm with
Enter
.