![MKS 670B Instruction Manual Download Page 131](http://html1.mh-extra.com/html/mks/670b/670b_instruction-manual_1813426131.webp)
MKS Device Dependent Messages
Chapter Seven: IEEE-488 Communications Option
116
Channel-Specific Messages (Continued)
Function
Message
Enable Heater
Command
Shortcut command
Query
Shortcut query
:route:control:heater(@chan #) State
:rout:cont:heat(@chan #) State
:route:control:heater(@chan #)?
:rout:cont:heat(@chan #)?
where chan # equals:
1 for Channel 1
2 for Channel 2
3 for Channel 3
where State equals:
ON for On or enabled
OFF for Off or disabled
Report Heater Current
Command
Shortcut command
:measure:heater (@chan #)
:meas:heat (@chan #)
where chan # equals:
1 for Channel 1
2 for Channel 2
3 for Channel 3
The 670 instrument will report the percentage of power
used by the heater. (It does not require a query message.)
Temperature Compensation
Command
Shortcut command
Query
Shortcut query
:calculate:scan(@chan #):temcomp value
:calc:scan(@chan #):temc value
:calculate:scan(@chan #):temcomp?
:calc:scan(@chan #):temc?
where chan # equals:
1 for Channel 1
2 for Channel 2
3 for Channel 3
value is an ASCII number between 0 and 999, inclusive
(value of 500 disables the feature)
Table 33: IEEE-488 Channel-Specific Messages
(Continued on next page)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com