![MKS 20340-341-T1 Instruction Manual Download Page 26](http://html1.mh-extra.com/html/mks/20340-341-t1/20340-341-t1_instruction-manual_1813378026.webp)
3 Installation
26
Series 340 Vacuum Gauge Controller
Instruction Manual - 340005 - Rev. D
3.3.3 Compression Mount (Quick Connect)
Do not use this type of fitting for positive pressure applications.
The gauge tube port is designed to fit a standard 1/2 in. compression (quick connect) mount such as
the Cajon Co.
®
Ultra-Torr
®
fittings.
Remove the cap plug from the gauge tube port, insert the gauge tube port into the compression
fitting and finger tighten the press ring. If a seal is not achieved it may be due to extreme cleanliness
of the O-ring. A light film of vacuum grease such as Apiezon will insure sealing and is normally
preferable to the use of pliers or pipe wrench to further tighten the press ring. You may point the
electrical pins of the gauge tube anywhere you wish in a 360 degree horizontal circle for optimum
routing of the gauge tube cable.
3.3.4 1/8 NPT Mount
The threads on the gauge tube port will fit a standard 1/8 NPT female fitting. Wrap the threads of the
gauge tube port with Teflon
®
tape and screw these threads into the system fitting hand tight. Do not
use any wrench or tool. The gauge tube body functions adequately as its own wrench. Tighten only
sufficiently to achieve a seal. When the threads have been tightened to the point where a seal is just
achieved, about one-half turn additional tightening is all that can be gained without overstressing
the tube port.
3.3.5 NW16KF Flange Mount
The KF mounting system requires an O-ring and centering ring to be placed between the mating
flanges. The flanges are then held together with the aluminum flange clamp by tightening the wing
nut. Maximum pressure for this style mounting system is 1000 Torr absolute.
Figure 3-7
Convectron Gauge Mounting Configurations