![Mizuno MOT-VS600 Operator'S Manual Download Page 40](http://html1.mh-extra.com/html/mizuno/mot-vs600/mot-vs600_operators-manual_1812848040.webp)
Lateral tilt
36
4
Operation
Left down
1. Press
and
simultaneously.
The tabletop will move to the left down position
as viewed from the head end.
REV. TREND.
TREND.
TILT LEFT
TILT RIGHT
BACK UP
BACK DOWN
TABLE UP
TABLE DOWN
SLIDE HEAD
BRAKE UNLOCK
LEVEL
ENERGIZE
ON/OFF
PRESS TO
LOCK BRAKES
MOT-VS600DK
REFLEX
FLEX
SLIDE FOOT
KIDNEY DOWN
KIDNEY UP
NOTE
- The maximum angle achieved
in the left down position is 35°
from the level position.
- When lateral tilt is operated in
the opposite direction from the
current lateral tilt position, it will
stop in the level position. Once
it stops, release the switch and
press it again, it will continue
tilting toward the other side.
- When the height from the floor
to the table top is in the range
of 720 mm (28.35 in) to 620
mm (24.41 in), the maximum
angle of the left down position
is 15°(MOT-VS600DK/MOT-
VS600DHK) or 30°(MOT-
VS600D/MOT-VS600DH).
*1 MOT-VS600DK/MOT-VS600DHK
*2 MOT-VS600D/MOT-VS600DH
- When the height from the floor
to the table top is 620 mm
(24.41 in) or less, the maximum
angle of the left down position
is 10°.
35
°
35
°
WARNING
Make sure the patient on the
table is secured when tilting
the table. Otherwise the patient
may fall off the table.
Left
down
15°(*1)
30°(*2)
15°(*1)
30°(*2)
620mm (24.41 in) to
720mm (28.35 in)
10°
10°
620mm (24.41
in) or less
Special notice
- About the operating table for movable range
cancelling specifications
If the same operation is performed again after it is
momentarily paused by the movable range that is
described in the NOTE field, it will be operated up
to the maximum movable range while the buzzer is
sounding.
Since the tabletop may bump against the floor or
base, and the slide cylinder may bump against the
column, so pay careful attention when operating it.
This covers the operating table applied with the “
●
”
label which is described in “Other labels” (P. 12).
- About the operating table for movable range setting
specifications
The operating table which is not applied with the “
●
”
label has movable range settings, and it will operate
as per the descriptions in the NOTE field.
Summary of Contents for MOT-VS600
Page 2: ......
Page 4: ......
Page 5: ...1 Introduction 1 Introduction ...
Page 6: ...2 1 Introduction ...
Page 10: ...6 1 Introduction ...
Page 11: ...2 Before use 2 Before use ...
Page 19: ...3 Component identification 3 Component identification ...
Page 24: ...20 3 Component identification ...
Page 61: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Page 64: ...60 5 Maintenance and inspection ...
Page 65: ...6 Specifications 6 Specifications ...
Page 69: ...7 Troubleshooting 7 Troubleshooting ...
Page 71: ...8 Before contacting for repairs 8 Before contacting for repairs ...