SPECIAL USER ATTENTION
NOTE
x
The Emergency Brake Release Valve
must be closed and tightened (coun-
ter-clockwise) before activating any hy-
draulic function.
x
If the Emergency Brake Release Valve
has been operated, the BRAKE
UNLOCK button on the pendant control
will have to be pressed before brakes
will lock again.
NOTE
To make the Back Section easier to
handle, remove the Head Section and
X-ray Top prior to removing the Back
Section.
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
NOTE
Normal table top position is with the
head (and back) section over the power
cord end of the base.
WARNING
Always lock the table top in position af-
ter rotation. DO NOT rotate the top with
an unevenly distributed patient weight
load as instability may result.
WARNING
x
Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes
are set before transferring the patient.
x
Exercise caution with the table top ro-
tated 90° to the base since an improp-
erly distributed patient load may cause
the table to be tipped over.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and
cardiac defibrillator monitors. Improper
operation procedures might cause a
burn on an operators or patient, and
also cause the malfunction of the
product.
WARNING
x
When a patient is put on a mattress and
his/her position is set up, a patient shall
be set up to more than 1 cm apart from
a side rail so that a patient dose not
touch on side rails (metal).
x
The mattress should be attached in
order not to come off while being in use.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
WARNING
The antistatic properties of the table are
dependent on the use of the original
pad set which was furnished with the
table or an alternate approved re-
placement.