2) Installation procedure 

Install the U-WAVE-R at a location with a good view and eliminate any obstacle between the 

U-WAVE-R and the U-WAVE-T. If installing it on a wall, etc., use the installation board   

(No. 02AZD815) from among the standard accessories. To install the U-WAVE-R on a wall, 

observe the following procedure. 

(1) Attach the mounting plate to the U-WAVE-R backside with cross-recessed head tapping 

screws (nominal size 2.6 × 6 No. A131-6221CP 2pcs.) in the standard accessories. 
Screw holes are provided on the U-WAVE-R backside so that the mounting plate can be 
fixed at one of three positions. Attach the mounting plate at an appropriate position   
where the U-WAVE-R is easily mounted. 

(2) Install the U-WAVE-R on a wall at a desired location with the two screws. 

 When installing the U-WAVE-R, refer to the following figure. Carefully install the U-WAVE-R 

so that it will not drop from the wall. 

 
• Figure of the installation board

(Unit : mm)

 

 

• Figure of how to attach the plate to the U-WAVE-R

 

(Attachable at one of three positions.) 

 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 

 

  [4] Functions 

1) Communication Function 

 

The U-WAVE-R displays the data received from the U-WAVE-T. 

Observe the following procedure for output operation. 

(1) Connect the U-WAVE-R to a personal computer with a USB cable. 
(2) Boot the U-WAVEPAK. 
(3) Please read the 

“U-WAVE Quick manual “ and “U-WAVEPAK User’s Manual“ in   

“PDF_Manual” folder of the CD supplied as a standard accessory for setting correctly. 

(4) Press the DATA switch on the connecting cable connected to the U-WAVE-T or the DATA 

switch on a Measuring tool to transmit data. 

(5) The U-WAVE-R displays the received data and wireless transmits the confirm signal to   

the U-WAVE-T. 

(6) U-WAVE-T receives the confirmation signal from U-WAVE-R, and can confirm the success   

(or failure) with LED (Buzzer). 

 

IMPORTANT 

•  Log in to Windows by ‘Power Users’ or above. 
•  The U-WAVE-PAK automatically transmits the measurement data to the active application 
    software (such as Microsoft Excel or Microsoft word). Before booting multiple application 
    software simultaneously, be sure to activate the application software that you want to use. 
•  U-WAVE-R receives the data from the U-WAVE-T and wireless transmits the confirmation   

signal to the U-WAVE-T, even if U-WAVEPAK doesn't start . 
Please confirm measurements are displayed in the application software that is before the   
measurement begins that wants to be used. 

 

2) Low power supply alarm 

Red LED blinks when the voltage of the power supply from the personal computer decreases. 
(Green LED is in the state of lighting.) 
Please change to a new USB cable at such a state. 

 

3) Initializing the setting information 

If communication is disabled while using the U-WAVE-R, first refer to section 5.   
If communication is still disabled, initialize the information set in section 3 to default settings, 
and then retry setup. 
Initialize the setting information with the following procedure. 

(1) Close U-WAVEPAK activated on the personal computer, and then pull off the cable of the 

  U-WAVE-R from the USB cable on the computer. 

(2) While holding down the INIT. switch on the U-WAVE-R, plug the cable from the U-WAVE-R   

into the USB cable on the personal computer again. Wait for 3 seconds or more in this state. 
The setting information is initialized. At this time check that the green LED is blinking. 

(3) Implement the setup again. 

 

IMPORTANT 

Once initialization is performed, the setting information used until then is all cleared. 

 

[5] Troubleshooting 

If any trouble is not corrected even when the following actions are taken, contact a Mitutoyo 
service center. 

Please read the final page of 

“U-WAVEPAK User’s Manual“ in “PDF_Manual” folder of   

the CD supplied as a standard accessory about Mitutoyo service center. 

 

1) Confirmation of U-WAVE-R

 

(1) The green LED will not light up (or start blinking). 

•  Is the power to the personal computer turned on? 
•  Is the USB cable connected between the U-WAVE-R and personal computer? 

The U-WAVE-R is supplied with power from the personal computer, and thereby no power   
will be supplied to the U-WAVE-R unless the personal computer is turned on. Check the   
USB cable for proper connection. 

•  Is the USB driver installed? 
 

(2) The red LED is blinking with the green LED being lit. 

•  Is any undue force applied to the USB cable? 

Check the connecting condition of the USB cable. Also, the USB cable might be broken 
internally. Check it out by using a new USB cable. 

 

(3) The green LED and red LED keep blinking alternately. 

•  Has setup been completed? 

Referring to the U-

WAVEPAK User’s Manual, set up the system correctly. 

•  Does any U-WAVE-R with the same group ID and band ID exist near the U-WAVE-R?   

Set different information for each ID in the U-WAVE-R. 

 

 
 
 
 

2) Confirmation of U-WAVE-T

 

(1) Data transmission always fails and the red LED lights up. 

•  Is the LCD of the Measuring tool put in the count display state? 

Communication is not possible if the LCD on the connected Measuring tool is turned off or in 

the data entry state. Put the Measuring tool LCD in the count display state. 

•  Has the battery voltage come down? 

If the battery is consumed to a low voltage, the red LED is blinking and data transmission is 

disabled. 

Replace the battery with a new one. 

•  Is the U-WAVE-T installed correctly on the Measuring tool? 

Recheck that the U-WAVE-T and the Measuring tool are properly mated using the supplied 

connecting cable. 

Also, when using the specific connecting cable (02AZD791A, B), be sure to mate the 

black-marked side of the connecting cable with the U-WAVE-T. 

•  Is the device ID described in the label on the U-WAVE-T set in the connecting state on the 

U-WAVEPAK? 
Unless it is in the connecting state, communication is not implemented. 

Referring to read the 

“U-WAVE Quick manual “ and “U-WAVEPAK User’s Manual“ in   

“PDF_Manual” folder of the CD supplied as a standard accessory, set up the system 
correctly. 

 

(2) The U-WAVE-R search cannot be executed, blinking orange color LED disappears, 

and red LED blinks. 

•  Is the U-WAVE-T installed correctly on the Measuring tool? 

Recheck that the U-WAVE-T and the Measuring tool are properly mated using the supplied 

connecting cable. 

      If U-WAVE-T is installed correctly on the Measuring tool correctly, initialize the setting 

  information and 

then retry communication after referring “section 6 (2) in U-WAVE-T manual” 

 

3) Confirmation of Wireless Communication Environments

 

Data communication is not stabilized, causing an error frequently. 

•  Is the U-WAVE-R separated too far from the U-WAVE-T? 
•  Is there any obstacle between the U-WAVE-R and the U-WAVE-T? 

First, bring these instruments close to each other, and then try communication between 

them. If an obstacle such as a wall or a metallic partition exists between the U-WAVE-R and 
U-WAVE-T, communication may be affected adversely. 

Also, the communication condition may be improved by changing the U-WAVE-R location and 

orientation. 

•  Is any microwave oven, wireless LAN, or Bluetooth used nearby? 

A wireless communication device or electric household appliance using a frequency of 2.4 

GHz, which is close to the frequency used in the U-WAVE-R, might be affected adversely. 
Particularly, medical equipment could cause a threat to life due to electromagnetic 
interference. Carefully use the U-WAVE-R while separating sufficiently from such 
equipment. 

Also, check the situation of used bands at the periphery with the U-WAVEPAK and change 

the current band to that with better communication quality. It may be possible to stabilize 
communication. 

Make an attempt on this, referring to 

“U-WAVEPAK User’s Manual“ in “PDF_Manual” folder of 

the CD supplied as a standard accessory. 

 

[6] Specifications 

Country 

Japan, EU member countries, U.S.A, Canada 

Code No. 

02AZD810D 

Certification number 

005WWCA0167 

Japan

 

VXU-02AZD810D (U.S.A)   

4396B-02AZD810D (Canada) 

Transmission output 

Less than 0 dBm (1 mW)

 

Distance of communication 

Approx.20 m 

(at line-of-sight distance under office environments) 

A Conformity standard 

Japan

    ARIB STD-T66 

Europe   

RE directive   

U.S.A

      47 CFR Part 15.247:(Subpart :C)   

47 CFR Part 15,(Subpart :B) 

Canada 

  RSS-210 (Issue 7) and RSS-Gen (Issue 2)

   

ICES 003 (Issue 4) 

Wireless communication protocol 

IEEE802.15.4 compatible 

Communication frequency 

2.405 to 2.475 GHz 

Used band 

15 channels (at intervals of 5 MHz) 

Modulation method 

DSSS (Direct Sequence Spread Spectrum) 

Wireless communication speed 

250 Kbps 

LED display 

Green/red 

Output to a personal computer 

USB2.0 12 Mbps (equivalent to USB1.1) 

Power supply system 

USB bus power system 

Operating temperature (humidity) 

0 to 40°C (20 to 80%RH, with no condensation) 

Storage temperature (humidity) 

-10 to 60°C (20 to 80%RH, with no condensation) 

External dimensions 

140 × 80 × 31.6 (mm) 

U-WAVE-R mass 

Approx. 130 g 

 

○ Standard accessories 

  U-WAVE-R 

User’s Manual 

this manual

  No. 99MAL109B 

  U-WAVE QUICK Manual No. 99MAL110B 

  USB cable (1m) No. C177-008 

  An Installation board No. 02AZD815 

  Cross-recessed head tapping screws   

nominal size 2.6 × 6 No. A131-6221CP 2pcs. 

U-WAVEPAK No.02ARB110

 

(The following contents bundle into the CD) 

 

  U-WAVEPAK Program 

 

  USB Device Driver 

 

  Data Collection Macro

 

for U-WAVE 

 

  U-WAVEPAK User

’s Manual 

 

No. 99MAL216 

  Warranty card 

   

 

Declaration of conformity document (copy) 

 

Holes for attaching to the U-WAVE-R (2 pieces) 

Holes for mounting on a wall (2 pieces) 

   