Table 3
Impact device
DC(D)/DL
D+15
C
G
E
Impacting energy
Mass of impact body
11mJ
5.5g/7.2g
11mJ
7.8g
2.7mJ
3.0g
90mJ
20.0g
11mJ
5.5g
Tip hardness:
Dia. tip:
Material of tip:
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten
carbide
1600HV
3mm
Tungsten
carbide
1600HV
5mm
Tungsten
carbide
5000HV
3mm
Synthetic
diamond
Impact device diameter:
Impact device length:
Impact device weight:
20mm
86(147)/
75mm 50g
20mm
162mm
80g
20mm
141mm
75g
30mm
254mm
250g
20mm
155mm
80g
Max. hardness of sample
940HV
940HV
1000HV
650HB
1200HV
Mean roughness value of
sample surface Ra:
1.6μm
1.6μm
0.4μm
6.3μm
1.6μm
Min. weight of sample:
Measure directly
Need support firmly
Need coupling tightly
>5kg
2
~
5kg
0.05
~
2kg
>5kg
2
~
5kg
0.05
~
2kg
>1.5kg
0.5
~
1.5kg
0.02
~
0.5kg
>15kg
5
~
15kg
0.5
~
5kg
>5kg
2
~
5kg
0.05
~
2kg
Min. thickness of sample
Coupling tightly
Min. layer thickness for
surface hardening
5mm
≥0.8mm
5mm
≥0.8mm
1mm
≥0.2mm
10mm
≥1.2mm
5mm
≥0.8mm
Size of tip indentation
Hardnes
s
300HV
Indentation
diameter
Depth
of
indentation
0.54mm
24μm
0.54mm
24μm
0.38mm
12μm
1.03mm
53μm
0.54mm
24μm
Hardnes
s
600HV
Indentation
diameter
Depth
of
indentation
0.54mm
17μm
0.54mm
17μm
0.32mm
8μm
0.90mm
41μm
0.54mm
17μm
Hardnes
s
800HV
Indentation
diameter
Depth
of
indentation
0.35mm
10μm
0.35mm
10μm
0.30mm
7μm
--
--
0.35mm
10μm
Suitability
DC: Test hole
or
hollow
cylindrical;
DL:Testslende
r
narrow
groove or hole
Test
groove or
reentrant
surface
Test small,light,
thin parts and
surface
of
hardened layer
Test large,
thick,heav
y
and
rough
surface
steel
Test
super
high
hardness
material