The handpiece receptacle is keyed in order to facilitate connection. The red dot on top of the receptacle must be in line with
the corresponding red dot on the handpiece cable.
6.2.
Menu Functions
The standard screen is the Main Screen. Additional screens are the Options and the Help Screen. Both the Options and
Help screens can be accessed by pressing the menu button to toggle through the three main screens; Main Menu, Options
and Help.
Main Screen
The Main Screen allows control of the main system functions such as Amplitude, Pulse and Flow.
1
A B C D E F
Figure 6.2.1 Main screen
Amplitude Control
The amplitude can be set between 0 and 10. Press A to increase and B to decrease the amplitude in increments of 1.
The default setting for amplitude is 7. Refer to section 1.1 for further details on the Amplitude feature.
Pulse Control
The pulse can be set between 50% and 100%. Press C to increase and D to decrease the pulse in increments of 10%.
The default setting for pulse is 100%. Refer to section 6.3 for further details on the Pulse feature.
Flow Control
The flow can be set between 20% and 100%. Press E to increase and F to decrease the flow. The flow is increased and
decreased in increments of 5% from 20%-40% and increased and decreased in increments of 10% from 40% to 100%.
The default setting is 70%. Refer to section 6.3 for further details on the Flow feature.
1. Menu Button
A-F Custom buttons