Operation Manual
____________________Function of Controls and Indicators
_________________________________________________________________________________________________
5-3
XL2010 & XL2015 FUNCTIONS
POSITION
NO.
CONTROLS
FUNCTION
1.
Overload Indicator
Indicates a load condition greater than the maximum output
power of the generator or that the unit requires tuning or
other readjustments.
2.
Tuning Switch
Activates tuning circuit.
3.
Power Switch
ON Position (I) – energizes generator.
OFF Position (O) – de-energizes generator.
4.
Power Output Display
Indicates a
RELATIVE
percentage of maximum available
ultrasonic power delivered to the horn (i.e. 40% output = 0.4
x 550 Watts available = 220 Watts delivered).
5.
PULSAR
TM
Cycle Timer
(XL2015 Only)
Activates pulsed energy. Pulse width adjusts from 0.05 of a
second (5% duty) to 0.95 of a second (95% duty) every
second.
6.
Amplitude Control Knob
Controls the amplitude of ultrasonic vibration at the tip of
the horn.
7.
Tuning Control Knob
Allows adjustment of generator frequency to optimize
performance of convertor/horn assembly. Also used to
verify condition of horn and/or tip.
8.
SHV Cable Connector
Connects convertor cable to generator.
9.
Footswitch Jack
Connects external ON/OFF control (Foot Switch or Timer).
10.
Power Receptacle
Connects line cord to generator.
11.
Fuse (10 amp 115V)
( 8 amp 220V)
Protects against electrical overload.