Instructions For Use – neXus Ultrasonic Surgical Aspirator
Doc #: 100-10-1000 rev A
p 29
6.4.2 Irrigation & Aspiration Tubing Installation
Open the irrigation pump cover and seat the irrigation tubing so that it is centered to the pump rollers and aligned with
the “V” slot as shown. Close the pump cover taking care not to pinch you fingers in the closure. The LED light above the
pump cover will turn from blue to green and the blue water symbol box on the screen will be checked indicating a
correct connection followed by the Priming screen.
Place the irrigation tubing onto the rollers as shown below. The tubing shall lie in the “V” slots on the right and left side
of the pump housing. Once positioned, close the pump cover. The blue LED above the pump head shall change from
blue to green. A check mark shall appear on the screen next to the water drop icon followed by the Priming screen.
6.4.2.1 Irrigation Tubing Installation
CAUTION:
o
Do not place the soft silicone tube behind or in front of the rollers (latch removed in illustrations)
o
Do not pinch the soft silicone tube when the latch is locked.
o
Do not pinch barb fittings when closing the latch.
6.4.2.2 Aspiration Tubing Installation