Installing the NC-2000 Patient Unit
3.1
Overview
The NC-2000 Patient Unit is the backbone of the wireless mesh network that connects all the
Nurse Call and Emergency System devices together. End devices (for example, the Pull
Station and the ZF3) send messages to the Patient Unit, and the Patient Unit then passes
these messages to the NC-2100K Coordinator Bundle and the Nurse Call and Emergency
System Central server for processing.
Figure 10 shows the front of the Patient Unit and its features, which include:
•
Four LED status indicators
•
Speaker and microphone for two-way voice communication with an attendant
•
Ultrasound transmitter to locate portable MiCare end devices in the room
•
Eight buttons (Code Blue, Emergency, Cancel, Nurse In, Nurse Out, Staff Assist,
Maintenance, and Custom Programming)
•
Two bed cord jacks
Figure 10 The front of the NC-2000 Patient Unit
The back of the Patient Unit has terminals for the following connections:
Install in accordance with the Canadian Electrical Code or the National Electrical
Code, and comply with all local regulations. Final acceptance subject to the Local
Authority Having Jurisdiction (AHJ).
!
Cancel
Nurse In
Nurse Out
Staff Assist
Maintenance
Custom Programming
Emergency
Code Blue
Bed Cord 2
Bed Cord 1
LED Status Indicators
Microphone
Speaker
Ultrasound
Transmitter