5. PROBLEM SOLVING
PROBLEM
MESSAGE
POSSIBLE CAUSES
REMEDY
MiniSpir does not connect with
the PC
\
The USB cable is not correctly
connected
Check the correct connection of the
USB cable side Pc
\
The driver doesn’t work correctly Check the presence of the device in the
list of USB devices connected. Try to
remove and connect the device.
Spirometry data at the end of the
test are not acceptable
\
The turbine don’t rotate correctly
Clean the turbine and check another
time; use a new turbine
\
The test is performed in a wrong
way
Repeat the test following the
indications on the screen
During an oximetry test the
values are wrong, irregular
\
The sensor is positioned in a
wrong way or the perfusion index
of the patient is low
Put the sensor in another position.
\
The patient is moving
To obtain an accurate measurement
the patient should not make sudden
movements.
6. LIMITED WARRANTY CONDITIONS
MiniSpir
, together with its standard accessories is guaranteed for a period of 12 months if intended for professional use (doctors,
hospitals, etc.).
The warranty is effective from the date of purchase contained in the relevant sales invoice or proof of purchase.
The instrument must be checked at the time of purchase, or upon delivery, and any claims must be made immediately in writing
to the manufacturer.
This warranty covers the repair or the replacement (at the discretion of the manufacturer) of the product or of the defective parts
without charge for the parts or for the labour.
All batteries and other consumable parts, reusable turbine included, are specifically excluded from the terms of this guarantee.
This warranty is not valid, at the discretion of the manufacturer, in the following cases:
•
If the fault is due to an improper installation or operation of the machine, or if the installation does not conform to the current
safety norms in the country of installation.
•
If the product is utilised differently from the use described in the Users Manual.
•
If any alteration, adjustment, modification or repair has been carried out by personnel not authorised by MIR.
•
If the fault is caused by lack of or incorrect routine maintenance of the machine.
•
If the machine has been dropped, damaged or subjected to physical or electrical stress.
•
If the fault is caused by the mains or by another product to which the instrument has been connected.
•
If the serial number of the instrument is missing, tampered with and/or not clearly legible.
The repair or replacement described in this warranty is supplied for goods returned at the customer
s’ expense to our certified
service centres. For details of these centres please contact your local supplier of the spirometer or contact the manufacturer
directly.
The customer is responsible for the transportation and for all transport and customs charges as well as for delivery charges of the
goods both to and from the service centre.
Any instrument or accessory returned must be accompanied by a clear and detailed explanation of the defect or problem found.
If units are to be returned to the manufacturer then written or verbal permission must be received before any instruments are
returned to MIR.
MIR Medical International Research, reserves the right to modify the instrument if required, and a description of any modification
made will be sent along with the returned goods.