5.  PROBLEM SOLVING 

 

PROBLEM 

MESSAGE 

POSSIBLE CAUSES 

REMEDY 

MiniSpir  does  not  connect  with 
the PC 

 

The USB cable is not correctly 
connected 

Check the correct connection of the 
USB cable side Pc  

The driver doesn’t work correctly  Check the presence of the device in the 

list of USB devices connected. Try to 
remove and connect the device. 

Spirometry data at the end of the 
test are not acceptable 

The turbine don’t rotate correctly 

Clean the turbine and check another 
time; use a new turbine 

 

The test is performed in a wrong 
way 

Repeat the test following the 
indications on the screen 

During  an  oximetry  test  the 
values are wrong, irregular  

The sensor is positioned in a 
wrong way or the perfusion index 
of the patient is low 

Put the sensor in another position.  

 

The patient is moving 

To obtain an accurate measurement 
the patient should not make sudden 
movements. 

 

6.  LIMITED WARRANTY CONDITIONS 

 

MiniSpir

, together with its standard accessories is guaranteed for a period of 12 months if intended for professional use (doctors, 

hospitals, etc.). 
The warranty is effective from the date of purchase contained in the relevant sales invoice or proof of purchase. 
 
The instrument must be checked at the time of purchase, or upon delivery, and any claims must be made immediately in writing 
to the manufacturer. 
 
This warranty covers the repair or the replacement (at the discretion of the manufacturer) of the product or of the defective parts 
without charge for the parts or for the labour. 
 
All batteries and other consumable parts, reusable turbine included, are specifically excluded from the terms of this guarantee. 
 
This warranty is not valid, at the discretion of the manufacturer, in the following cases: 
 

 

If the fault is due to an improper installation or operation of the machine, or if the installation does not conform to the current 
safety norms in the country of installation. 

 

If the product is utilised differently from the use described in the Users Manual. 

 

If any alteration, adjustment, modification or repair has been carried out by personnel not authorised by MIR. 

 

If the fault is caused by lack of or incorrect routine maintenance of the machine. 

 

If the machine has been dropped, damaged or subjected to physical or electrical stress. 

 

If the fault is caused by the mains or by another product to which the instrument has been connected. 

 

If the serial number of the instrument is missing, tampered with and/or not clearly legible. 

 

The repair or replacement described in this warranty is supplied for goods returned at the customer

s’ expense to our certified 

service  centres.  For  details  of  these  centres  please  contact  your  local  supplier  of  the  spirometer  or  contact  the  manufacturer 
directly. 
 
The customer is responsible for the transportation and for all transport and customs charges as well as for delivery charges of the 
goods both to and from the service centre. 
 
Any instrument or accessory returned must be accompanied by a clear and detailed explanation of the defect or problem found. 
If  units  are  to  be  returned  to  the manufacturer  then  written or  verbal  permission  must  be received  before  any instruments are 
returned to MIR. 
 
MIR Medical International Research, reserves the right to modify the instrument if required, and a description of any modification 
made will be sent along with the returned goods. 
 

Summary of Contents for MiniSpir Light

Page 1: ...MiniSpir User manual MINISPIR User manual Rev 1 9 page 1 of 18 User Manual Rev 1 9 Issued on 09 10 2020 Approved on 09 10 2020...

Page 2: ...the PC The device can then be used with the winspiroPRO software Keep the original packaging In the event that your device requires attention then always use the original packaging to return it to th...

Page 3: ...n electromagnetic environments 7 1 3 Unforeseen errors 8 1 4 Labels and symbols 8 1 4 1 Identification label 8 1 4 2 ESD Electrostatic discharge sensitivity symbol 9 1 5 Product description 9 1 6 Tech...

Page 4: ...e for ensuring that the device is stored and used in appropriate environmental conditions as specified in paragraph 1 6 3 WARNING If the device is exposed to unsuitable environmental conditions this c...

Page 5: ...ned before each spirometry test and a new disposable mouthpiece must always be used for each subject The use of an anti bacterial filter is at the discretion of the doctor If a disposable turbine is u...

Page 6: ...rect functioning of the device and thus of incorrect test results and create inconvenience to the patient The user is responsible for obtaining the correct type of mouthpieces for the device Those req...

Page 7: ...igurable parameters should only be made by qualified personnel However an incorrect set up of the parameters does not put the patient at risk When connected to other devices to preserve the safety oft...

Page 8: ...nnect to other devices such as PC or printer Use only the USB cable supplied by the manufacturer and observe the safety regulations of IEC 60601 1 1 SpO2 Warning symbol for the SpO2 port for oximetry...

Page 9: ...ope used for work instructions 600 7 000 volts Worker at a bench 700 6 000 volts If two items are at different electrostatic charge levels as they approach one another a spark or Electrostatic Dischar...

Page 10: ...from healthy subjects MiniSpir is connected to a PC trough a USB port Data measured by MiniSpir are transferred to the PC in real time The Windows winspiroPro software allows to view the spirometric...

Page 11: ...e mL FIVC Forced inspiratory volume L FIV1 Volume inspired in the 1st second of the test L FIV1 FIVC FIV1 FIVC x 100 PIF Peak inspiratory flow L s FIF25 Maximum flow at 25 of FIVC L s FIF50 Maximum fl...

Page 12: ...he analysis period BPM Recording time Total time measure of SpO2 hh mm ss Analysis Time Total time of analysis hh mm ss T90 Time passed with SpO2 90 hh mm ss T89 Time passed with SpO2 89 hh mm ss T88...

Page 13: ...liant with EN 60601 1 2005 A1 2012 Error of displayed numeric value Flow measurement percentage error 5 Oximetry parameters measuring with accuracy defined in table on 1 6 2 Emission limits CISPR 11 G...

Page 14: ...struction or restriction plus one level of normal spirometry as shown in the following table normal mild moderate moderately severe severe very severe Through an analysis applied to some of the indice...

Page 15: ...de by the individual doctor Re usable finger sensor This sensor is recommended for patients weighing 20 Kg with limited activity Carry out an oximetry test as follows Connect the sensor to the instrum...

Page 16: ...line runs across the mid axis of the fingertip Wrap the bottom adhesive around the digit being careful not to cover the nail Fold the sensor s top over the digit making sure the light source is direct...

Page 17: ...he turbine such as threads or hair Any such deposit could brake or block the rotation of the turbine blade and thus compromise the measurement accuracy To clean the reusable turbine remove it from its...

Page 18: ...le turbine included are specifically excluded from the terms of this guarantee This warranty is not valid at the discretion of the manufacturer in the following cases If the fault is due to an imprope...

Reviews: