MIPEX-04 ULTRA-LOW POWER INFRARED GAS SENSOR
ESAT.100000.00.02 UM
Revision 02
December 12th, 2016
Page 19 of 41
THE SOLE PROPERTY OF OPTOSENSE LLC. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
Table 6. Basic sensor readings variability
6
Calibration
gas
Readings variability within a
temperature range
Additional variability
due to pressure
Additional variability
due to humidity
CH
4
±0.1% vol. or ±5% of indication
(whichever is greater) within the
range of +20…+25 °C
±0.2% vol. or ±30% of
indication
(whichever
is greater) at 100 kPa
(test: 80 kPa, 100 kPa,
120 kPa)
±0.2% vol. or ±15% of
indication at 40 °C
(test: 20% RH, 50%
RH, 90% RH)
±0.2% vol. or ±10% of indication
(whichever is greater) within the
range of -
10…+20 °C and
+25…+40 °C
±0.4% vol. or ±20% of indication
(whichever is greater) within the
range of
-
40…-10 °C and +40…+60 °C
C
3
H
8
±0.05% vol. or ±5% of indication
(whichever is greater) within the
range of
+20…+25 °C
±0.1% vol. or ±30% of
indication
(whichever
is greater) at 100 kPa
(test: 80 kPa, 100 kPa,
120 kPa)
±0.1% vol. or ±15% of
indication (whichever
is greater)
at 40 °C
(test: 20% RH, 50%
RH, 90% RH)
±0.1% vol. or ±10% of indication
(whichever is greater) within the
range of
-
10…+20 °C and +25…+40 °C
±0.2% vol. or ±20% of indication
(whichever is greater) within the
range of -
40…-10 °C and +40…+60
°C
6
Table shows basic variability of MIPEX-04 readings. For each sensor modification, readings
variability stated in Table 6 is provided within a temperature range determined by a sensor RX code (see
Table 5).