MingTai SUN12 Operation Manual Download Page 4

 

 

person and equipment.

 

 

The  chloride  ion  is  an  important  factor  of  stainless  steel  corrosion  da

mage.  If  use  this  sterilizer  sterilization  items  containing  chlorine  ion,  must 
rinse  sterilization  indoor  wall  every  day

  in  order  to  avoid  deposition  on  i

nternal  stainless  steel  corrosion  of  chlorine  ion,  delayed  the  service  life  of 
the  product.

 

 

The product is only suitable for high temperature resistance, high humidity 

medical appliance sterilization box

  cannot be used with vaseline oil and powder 

of sterilization, etc.

 

 

3

Tech. Parameter 

Chamber size of the Steam Sterilizer(Diameter X Depth) 8L

 170mmX310mm 

 

Chamber size of the Steam Sterilizer (Diameter X Depth) 12L

200mmX360mm   

Chamber size of the Steam Sterilizer (Diameter X Depth) 16L

230mmX360mm 

Chamber size of the Steam Sterilizer (Diameter X Depth) 18L

249mmX355mm   

Chamber size of the Steam Sterilizer (Diameter X Depth) 23L

249mmX460mm 

Rated Voltage

………………………………………

230

, 50Hz

 

 

Rated Power

………………………………………

1800VA

           

2300VA 

Sterilizing (Pressure/temp.)

…………………………………

100-120KPa/121

º

                                                                                            200-220KPa/134

º

  (Both  of  above  are  Under  Standard  Atmospheric  pressure) 

Fuse-Melting pipe

………………………

.. 

……

……

.. 

…………

.. 250V/T 15A   

Capacity of water tank

………………………………………………………

2L 

Probable water consumption of one cycle 

……………………………

.. . .0.4L 

Environment temperature of operate

……………………………………

0-40

º

Type of sterilization cycles

………………………………………………

Class B 

Wire heat resistance level

…………………………………………………

70

º

N·W

    23L

…………………………………………………………………

53kg 

N·W

    18L

…………………………………………………………………

48kg 

N·W

    16L

…………………………………………………………………

47kg 

N·W

    12L

…………………………………………………………………

46kg 

N·W

    8L

…………………………………………………………………

 44kg 

 

 

 

 

 

 

 

Summary of Contents for SUN12

Page 1: ...ration Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 M Steam Sterilizer File No MTQM C1 7...

Page 2: ...3 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 6 8 Emergency Condition 9 9 Maintenance and Service 11 10 Transportation and Storage 12 11 Attention 13 Appendix NO 1 The wor...

Page 3: ...eliable product in sterilizing 1 Printer 2 USB 3 Used water outlet 4 Used water outlet 5 Power switch 6 Power source socket 7 Fuse 2 Application Scope This machine can be applied in sterilizing for th...

Page 4: ...erilizer Diameter X Depth 8L 170mmX310mm Chamber size of the Steam Sterilizer Diameter X Depth 12L 200mmX360mm Chamber size of the Steam Sterilizer Diameter X Depth 16L 230mmX360mm Chamber size of the...

Page 5: ...ACTURE AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY SERIAL NUMBER CONSULT INSTRUCTIONS FOR USE Earthing 5 Installation 5 1 Remove the packing Dismantle thin film open the packing get the equipm...

Page 6: ...d the around articles shall be 10cm as retained and the top surface shall be 20cm retained Used to exhaust or drain 5 4 The 2 front feet of Steam Sterilizer can be adjusted the height please ensure th...

Page 7: ...nz Zmax 0 114 OHM sind solche St rungen nicht zu erwarten Bitte kontaktieren Sie Ihre lokales Energieversorgungsunternehmen f r weitere Informationen PREPARATION BEFORE OPERATION Before operation plea...

Page 8: ...laying the air pressure inside the chamber Unit KPa 4 Door lock indicator light It is the light for door closing well can press START to working or during the working cycle the door were not opened 5...

Page 9: ...ture window display sterilizing temperature of selected program Pressure window display sterilizing pressure of selected program The Status window display selected cycle code as below 1 134 UNWRAPPED...

Page 10: ...ocess open the vacuum selection valve exhaust valve normally on and start the vacuum pump 300 seconds later Close the vacuum selection valve exhaust valve normally on and the vacuum pump to stop pumpi...

Page 11: ...3 1 3 Sterilizing time 5 5 20 20 25 Drying time 15 18 20 20 20 7 4 Emergency exit Long press the start key 5 s Machine will out of sterilization procedures meanwhile EC alarm 7 5 After sterilization u...

Page 12: ...sing slowly Inspect the filter inside the chamber keep it clean and no jam 6 E6 long du Problem of door switch or door being opened when working Inspect door switch or if door handle were revolved to...

Page 13: ...ooth cloth with distilled water to wipe the surface of the sealing ring or sealing cap gently The user shall discharge the sealing ring for further cleaning or replacement if the air leaking problem c...

Page 14: ...screwdriver when pull Do not use finger directly Operator also need ke ep away from it prevent burns Replace the safety valve This operation is limited to the professionals 1 Loosen the set screws of...

Page 15: ...ting tank shall be emptied once the water storing tank emptied 6 The door handle must be buttoned up before working 7 Don t open the door of Steam Sterilizer before seeing the 0 kPa displayed on the p...

Page 16: ...nge of temperature and pressure may cause the liquid bottle burst so that endanger the safety of the person and equipment 22 The chloride ion is an important factor of stainless steel corrosion damage...

Page 17: ...ing the working pressure pressure data In this status the chamber have pressure 4 Displaying the working time Sterilizing status In this status the working time and the count down figure are displayin...

Page 18: ...16 NO 2 Steam loop Work Description...

Page 19: ...ystem close the door then press START button to start Ningbo Mingtai Medical Instrument Co Ltd Yunzhen Road Industrial Zone Yunlong Town Yinzhou Ningbo City China Tel 86 0574 88320096 Fax 86 0574 8832...

Reviews: