background image

13                 F1A4 Operator Instructor Notes and Syllabus 

 

 

 

 

Minelab Countermine Division 

                                                            Safety through Excellence 

 

Thirty one 

 

Control Box Operation. 

 

Have the students unpack the Control Box from the bag and follow the demonstration 

while looking at the Control Box. 

 

Explain the function of each Control Box switch and button.  Key points to note are: 

 

  The  Ground  Balance  Switch  must  be  held  in  Balance  position  throughout  the 

Ground Balance procedure. 

  There must be no metal under the Coil when conducting Ground Balance. 

  The detector must be held perfectly still when conducting Noise Cancel. 

  There must be no metallic objects near the Coil during Noise Cancel. 

  When the Battery Low alarm sounds, batteries should be immediately changed 

(demonstrate ‘Battery Low’ alarm with a set of weak batteries). 

  The  F1A4  uses  a  Threshold  Tone  that  provides  a  very  soft  tone  to  provide  a 

positive  indication  to  the  operator  that  the  detector  is  working  correctly  –  the 

Threshold Tone also assists the operator in distinguishing tone changes caused 

by a small target. 

 

 

Thirty two 

 

Normal Procedures

 

Using  the  Field  Guide,  demonstrate  the  Normal  Procedures  (switch  on  and  Ground 

Balance). 

Key points: 

  During  the  six  seconds  start  up,  internal  diagnostic  checks  occur,  previous 

Ground  Balance  and  Noise  Cancel  settings  are  loaded  and  no  detection  is 

possible. 

  When conducting Ground Balance: 

o

  The Coil needs to be raised 10 cm from the ground. 

o

  Each complete up and down movement should occur over 3-4 seconds.  

o

  The Coil should be kept horizontal to the ground during the up and down 

movement. 

o

  Continue with the up and down movement until all ground interference is 

removed. 

o

  If the alleged ground interference does not reduce, it is likely that the Coil 

is  trying  to  Ground  Balance  a  piece  of  metal  –  select  another  area  to 

Ground Balance. 

o

  Once a Ground Balance procedure has been completed, the detector will 

remember  those  Ground  Balance  settings  even  after  it  is  switched  off 

and on again.

 

 

 

Thirty three 

 

Optional Procedures. 

 

Using  the  Field  guide  demonstrate  the  Optional  Noise  Cancel  and  Audio  Reset 

procedures: 

Key points: 

 

Noise Cancel  

o

  During  Noise  Cancel  it  is  important  to  keep  the  detector  still  –  during  this 

procedure  the  F1A4  electronics  scans  the  environment  over  4000 

frequencies  to  search  for  and  record  all  sources  of  interference  –  once 

Summary of Contents for F1A4

Page 1: ...OPERATOR INSTRUCTORS NOTES AND SYLLABUS F1A4 Metal Mine Detector Issue 1 st June 2005...

Page 2: ...n the course This can include some background information on the Minelab F1A4 Students will complete course registration administration as required If applicable students will be asked to stand and in...

Page 3: ...on before using the equipment If a fault develops an operator must be able to correctly name the faulty part of the detector Explain the importance of the Test Piece Four Equipment Specifications List...

Page 4: ...d is adjustable to various arm lengths Made of glass filled nylon and is very tough Identify the Locking Pin for connection of Control Box to the students Lower Shaft Explain that Lower Shaft cannot c...

Page 5: ...o Earset Speaker Off identifiable by a short length of green heatshrink on the cable shuts the detector Loudspeaker off when connected o Both Earset above are also available with an in built volume co...

Page 6: ...the stowed F1A4 components Mention the following Check for damage cleanliness Have students be able to name each item Stow away correctly Other Components Operations Manual waterproof paper Field guid...

Page 7: ...ly procedure in the following manner Secure the Handle As you take the F1A4 out of the Compact Bag secure the Handle in one place by locking the Handle Camlock This will prevent it from sliding up and...

Page 8: ...the following Control Box should be placed in the Compact bag when fitting removing batteries stops dirt and debris from getting on Control Box The batteries should be inserted from the centre outward...

Page 9: ...Dust Caps as required Test Piece being correctly stowed into the Control Box Bag Have the students take turns assembling and disassembling and watch for incorrect procedures Alternatively select one s...

Page 10: ...The major teaching points of this lesson are as follows Revision Principle of Operation Multi Period Sensing Demonstration of Ground Balancing TEACHING POINTS Revision Nominate selected students to as...

Page 11: ...a difficult concept to explain It is recommended that instructors make best use of PowerPoint presentation MPS produces a series of wide and narrow pulses that are transmitted into the ground from the...

Page 12: ...e target this will not happen if the target is a large piece of metal as the response to that will be louder Remove the target and pass only the mineralized rock under the Coil to show that the respon...

Page 13: ...or area suitable for the burial of targets confirm that the area is free of metallic contamination and large enough to accommodate the size of the class F1A4 detectors one per pair of students One set...

Page 14: ...ocedures Using the Field Guide demonstrate the Normal Procedures switch on and Ground Balance Key points During the six seconds start up internal diagnostic checks occur previous Ground Balance and No...

Page 15: ...e detector emits a loud oscillating tone when the charge in the batteries drops to a predefined low level There is no loss of sensitivity or capability before the LED flashes or the audible tone is he...

Page 16: ...etection and mapping against multiple and large targets Note that this will be demonstrated in the outdoor area but should first be explained in the classroom When large or multiple targets are found...

Page 17: ...turn it On Wait for the end of the start up tones and then perform the Noise Cancel procedure Wait 70 seconds for the Noise Cancel to end with the four beeps o Continue the above procedure if more th...

Page 18: ...erating procedures Forty six Care of the F1A4 Detector Discuss the points covered in the relevant PowerPoint slide Disassembly of the detector and its stowage in the Compact Bag is also part of this l...

Page 19: ...Countermine Division Safety through Excellence Forty eight QUESTIONS At the end of this period of instruction the student shall be able to carry out operator care and maintenance and correctly stow t...

Page 20: ...the following topics Ground Balance This is always impressive If the soil in the training area is benign prepare ahead of time and seek some mineralized soil from the local area or a local brick manu...

Page 21: ...he pitch of the tone From these changes in pitch you will quickly ascertain where you believe the centre of the target mass is Water Dig a hole and fill it with water or use a plastic bucket without a...

Page 22: ...at the end of these Instructor Notes Equipment Cleared evaluation area constructed as one mine lane 1m wide and 5m long Targets to bury in the test area as described below if inert mines are not avail...

Page 23: ...14 Here they must demonstrate that they can actually detect two targets Be careful to ensure that the students do not miss the smaller target by only recognizing a response from the larger target The...

Page 24: ...rea that they will investigate by prodding D One large target right on the edge of the lane To prove that not all mines are conveniently located in the middle of lanes E one small target single Test P...

Page 25: ...24 F1A4 Operator Instructor Notes and Syllabus Minelab Countermine Division Safety through Excellence This is to certify that DETECTOR MINELAB F1A4 METAL MINE MINELAB ELECTRONICS PTY LTD Signed Dated...

Reviews: