KT60558-1G
Sec 3: Mounting Procedure
䚷䚷
3-1. Typical mounting configuration
(1)
(2)
(3)
(4)
(5)
(1)
DY (dynamo)
(2)
Torque transducer TMHS-*NM
(3)
Coupling (H-flange)
(4)
Flexible coupling
(5)
Intermediate bearing
䚷䚷
3-2. Mounting flange
It may impede the power supply and cause zero balance shift or drifting
when metal objects' is close to the coil which attached to the
periphery of the flange on the equipment.
(1) Mounting unit outer diameter
Measured to the metal rim of the torque transducer flange.
䞉
Ø104 mm max for TMHS-100NM to 300NM
䞉
Ø111 mm max for TMHS-500NM and 1KNM
䞉
Ø153 mm max for TMHS-2KNM and 3KNM
(2) Distance from torque transducer flange edge face
䞉
At least 8 mm for A type mounting flange configuration (diameter extends to rotor)
䞉
At least 15 mm for B type mounting flange configuration (diameter extends to stator)
(3) Distance from torque transducer stator periphery (gap with protective cover, etc.)
䞉
At least 10 mm from antenna outer diameter
䐟
䐠
䐡
Mounting flange
Coil winding area
Rotor
Stator (antenna)
(1)
(2)
(3)
MinebeaMitsumi Inc. Sensing Device Business Unit