Bridge configuration
and circuit
Example
Bridge Box Wiring
Method
Remarks
Bridge Voltage
Output
One gage configuration
∙
Suited for where simple
tension, compression, or
bending force is applied
∙
Suited for where changes in
ambient temperature are
small.
∙
Calculated using the
calibration value as it is.
Bridge Voltage (E)
Output
One gage three wire configuration
∙
Suited for where simple
tension, compression, or
bending force is applied.
∙
Strain gage lead wires are
temperature compensated.
∙
Calculated using the
calibration value as it is.
Output
Bridge voltage (E)
One active gage, one dummy gage
configuration.
∙
Suited for where siple
tension, compression or
bending force is applied.
∙
Temperature compensation
using a dummy gage.
∙
Calculated using the
calibration value as it is.
Output
Bridge voltage (E)
Two active gage configuration
∙
Suited for where simple
tension, compression, or
bending force is applied.
∙
Temperature compensation
∙
Calculated using calibrated
value
x
1/(1 +
ν
)
∙
or signal value
x
1/(1 +
ν
)
Output
Bridge voltage (E)
Two active gage configuration
∙
Detects bending strain
∙
Eliminates tension and
compression strain.
∙
Temperature compensation
∙
Calculated using calibration
value
x
1/2 or siganal value
x
1/2
Output
Bridge voltage (E)
Opposite arm, two active gage
configuration
∙
Detect tension and
compression strain
∙
Eliminates bending strain
∙
Effects of changes in
temperature are doubled.
∙
Calculated using calibration
value
x
1/2 or signal value
x
1/2
Summary of Contents for DSA-631
Page 1: ...Dynamic Strain Amplifier DSA 631 INSTRUCTION MANUAL EN294 1320 E ...
Page 2: ......
Page 58: ...48 9 3 Outline dimensions 9 3 1 Single unit Unit mm ...
Page 59: ...49 9 3 2 Panel cut size Square hole Unit mm ...
Page 61: ...51 9 3 4 Rack mount case Rail for rack Long hole Unit mm ...
Page 62: ...52 9 3 5 Bridge box FA410 072 model Unit mm ...
Page 63: ......