MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-27
5.
Configuring Outputs
5.1.
Overview
Limit values can be configured to generate outputs at Out.1, Out.2, Out.3 ports. Several
signals allocated to output are linked by OR logic.
The device can generate the output based on
Process value (displayed value)
Measured value INP
Zero adjustment
5.2.
Configuring OUt.1 and OUt.2
Three signal limits are available with the device. Each limit value consists of a switch-on and a
switch-off point for definition of a hysteresis. The three pairs of values must be entered
according to the same principle. The limit values refer to the corresponding source.
This option can be used to obtain information on the filling status of a hopper and to
generate a signal. Each limit has three weight variables (lower limit, upper limit, and
hysteresis).
The procedure for configuring the outputs 1 and 2 via the device and tool are similar and
explained as below.