4.5 Ordering information
4.5.1
Load cells Inteco®/500 kg
75 t
Model
Max. capacity
E
max
Type
Inteco®/500 kg
500 kg
../D1/D1E
Inteco®/1 t
1 t
../D1/D1E
Inteco®/2 t
2 t
..D1/D1E/C3/C3E
Inteco®/3 t
3 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/5 t
5 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/10 t
10 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/20 t
20 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/30t
30 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/50 t
50 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/60 t
60 t
..D1/D1E/C3/C3E/C6/C6E
Inteco®/75 t
75 t
..D1/D1E/C3/C3E/C6/C6E
4.6 Technical data
4.6.1
Load cells PR 6203/500 kg
75 t
Designation
Description
Abbr. D1, D1E
C3, C3E
C6, C6E
Unit
Accuracy class
0.04
0.015
0.008
% E
max
Minimum dead load lowest limit of speciied
measuring range
E
min
0
% E
max
Maximum capacity
highest limit of speciied
measuring range
E
max
See Chapter
maximum load without
irreversible damage
up to E
max
= 50 t
E
lim
150
% E
max
E
max
= 60 t
E
lim
125
% E
max
Safe load limit
E
max
= 75 t
E
lim
100
% E
max
danger of mechanical de-
struction
up to E
max
= 50 t
E
d
300
% E
max
E
max
= 60 t
E
d
250
% E
max
Destructive load
E
max
= 75 t
E
d
200
% E
max
minimum load cell scale
interval, v
min
= E
max
/Y
Y
see following tables
Minimum LC verii-
cation
E
max
= 500 kg
Y
2500
…
…
Compression load cell Inteco® PR 6203
4 Speciications
EN-12
Minebea Intec