Selection is switched by tapping [NEG.] or [POS.] in a touch panel from OUT1 to OUT11.
[NEG.] is set as default.
10.4 Setting of servo control interface
1
PULSE F, PULSE R
PULSE F and PULSE R output the command pulse train that controls the servo
ampliier and the servo motor by the diferential line driver type. Please use them
after conirming the command pulse input form of the servo ampliier used. Refer
to Chapter
2 Encoder A-phase pulse and encoder B-phase pulse
Encoder A-phase pulse and encoder B-phase pulse can connect encoder A-phase
pulse and encoder B-phase pulse output from servo ampliier by diferential line
driver type.
3
Alarm signal
Please connect the alarm signal output from servo ampliier for alarm signal of this
unit. [SV.ERR ALM] lights when the alarm signal becomes efective. Refer to Chap-
ter
for switching the input efective state.
Note:
[SV.ERR ALM] lights when neither the connection nor the setting is correct.
4 Emergency stop signal
Please connect the switch of the b-contact for emergency stop signal. [SV.ERR
EMG] lights when the emergency stop signal is efective and the emergency stop of
the system with this unit becomes available. Refer to Chapter
when you do not
use the emergency stop signal.
Note:
[SV.ERR EMG] lights when not correctly connected.
5 Ready signal
Please connect the ready signal output from servo ampliier for ready signal of this
unit. [SV.ERR RD] lights when the ready signal becomes invalid. Refer to Chapter
for switching the input efective state.
Note:
[SV.ERR RD] lights when neither the connection nor the setting is correct.
Weight Controller CSD-918
10 I/O setting (I/O)
EN-90
Minebea Intec