7.10.5.1
Minimum load for application
The menu item [MIN.INIT] (3.6) sets the minimum load, i.e. the load that must be placed
on the weighing platform in order to carry out the application.
If the load on the platform is too light, the following will occur:
-
Error code [INF 29] appears,
-
The weighing platform is not initialized,
-
The preset reference sample quantity is saved.
Setting: [APPLIC.] - [APPLIC.1] - [PERCENT] - [MIN.INIT] (3.6) (see chapter
or
The minimum load can be set in 10 steps from 1 to 1000 digits (see available parameters).
The "digits" here refer to the scale intervals for the connected weighing platform. If the
interval of the connected platform is 1 g, for example, and 1000 digits are required, you
must place at least 1000 g (= 1000 intervals = 1000 digits) on the weighing platform for
initialization.
7.10.5.2
Resolution
The resolution [RESOLUT] (3.9) indicates the accuracy used to determine the reference
weight.
The default setting is [DISP.ACC] (3.9.1) "display resolution." The resolution is increased
when [10 FOLD] (3.9.2) "10-fold" or [100FOLD] (3.9.3) "100-fold" is selected. "10-fold"
increases the resolution of the net value by one step (display resolution x 10), "100-fold"
increases it two steps (display resolution x 100).
Setting: [APPLIC.] - [APPLIC.1] - [PERCENT] - [RESOLUT] (3.9) (see chapter
or
7.10.5.3
Decimal Places
In neutral measurement, not only whole numbers but also decimal numbers (for example,
1.25 o electrical cabling) can be displayed [DEC.PLCS] (3.10) "Decimal Places". The number
of decimal places displayed can range from none up to 3 places.
Setting: [APPLIC.] - [APPLIC.1] - [PERCENT] - [DEC.PLCS] (3.10) (see chapter
or
7.10.5.4
Parameter for Saving Weight Values
The weight on the platform is saved as a reference value as soon as the platform has
stabilized [SAVE WT] (3.11) "Parameter for saving weight values". "Stability" is deined as
the point at which the luctuation of a measured value lies within a deined tolerance. The
narrower the tolerance, the more stable the platform is at "stability".
The [ACC.STAB] (3.11.2) "increased stability" setting has a lower tolerance so that the
average piece weight saves is more accurate and the results more reproducible; however,
the measurement time can take longer.
Setting: [APPLIC.] - [APPLIC.1] - [PERCENT] - [SAVE WT] (3.11) (see chapter
or
1. Select menu [APPLIC.] - [APPLIC.1] - [PERCENT] with
key and open with
key (see chapter
2. To save the setting, press the
key.
3. To exit setup: Press the
key several times.