Care and Maintenance 9-9
NO.
Item
Method
Tool
1.
Casters
a) Visually check to confirm there is no any crack.
b) Operate the casters to confirm the locking and
releasing functions are normal.
none
2.
Connection of
the caster
a) Visually check to confirm that there is no skewness
and the connecting screws are free of breakage or
falling off.
b) Check with the spanner to make sure that there is no
looseness between the caster and the base
connection screw.
Inner
hexagon
wrench 8
3.
Handle
1. Visually check to confirm there is no any crack.
2. Rock the handle gently and then lift the ultrasound
machine to confirm that the handle is free of
looseness and it can accept normal force.
none
4.
Control panel
and turning
mechanism
Open the control panel to confirm that turning mechanism
is normal and the control panel is free of skewness and
looseness.
none
5.
Display and
turning
mechanism
a) Visually check to confirm there is no any crack to the
cover of display.
b) Manually operate the monitor to make sure that the
turning angle of display is normal and no looseness
exists.
none
6.
Other
mechanical
structures
Check to confirm that there is no looseness to other
mechanical parts, no crack to cover and no conductive
parts show in sight.
none
9.3.5
Electrical Safety
Inspection
Only technical professionals from Mindray or engineers authorized by Mindray after training can
perform electric safety inspection.
Please refer to appendix A: Electrical Safety Inspection for details.
Summary of Contents for Z6
Page 2: ......
Page 14: ......
Page 16: ...2 2 Product Specifications Left side 9 4 7 5 6 8 ...
Page 72: ......
Page 119: ...Structure and Assembly Disassembly 7 35 图 7 59 Disassembly of Speaker 8 ...
Page 120: ......
Page 132: ......
Page 142: ......
Page 152: ......
Page 168: ......
Page 171: ...Phantom Usage Illustration B 3 ...
Page 172: ......
Page 173: ...P N 046 010452 00 5 0 ...