D - 2
D.1.4
NIBP Physiological Alarm Messages
D.1.5
EWS Physiological Alarm Messages
Note:
N represents a numeric. XX represents RR, SpO
2
, BP-S, BP-D, BP-M, PR, Temp, EtCO
2
.
D.2
Technical Alarm Messages
This section lists technical alarms, their default priority, indication on alarm reset, and
the actions that can be taken when an alarm occurs.
Technical alarms give different alarm indicators when the alarm system is reset. In this
section we classify the technical alarms into three categories for easy clarification:
■
A: technical alarms are cleared. The monitor gives no alarm indications.
■
B: technical alarms are changed to the prompt messages.
■
C: the alarm is silenced and a
√
appears before the alarm message, indicating that
the alarm is acknowledged.
In the following tables we will use A, B, and C to refer to the indications on alarm reset.
Alarm message
Default
priority
Cause and solution
NIBP-S Extremely
High/NIBP-D
Extremely High/
NIBP-M Extremely
High
High
The NIBP value is higher than the NIBP Extreme
alarm high limit. Check the patient’s condition
and check if the alarm limit settings are correct.
NIBP-S Extremely
Low/NIBP-D
Extremely Low/
NIBP-M Extremely
Low
High
The NIBP value is lower than the NIBP Extreme
alarm low limit. Check the patient’s condition and
check if the alarm limit settings are correct.
Alarm message
Default
priority
Cause and solution
EWS Score > N
High/
Mediate
The total score exceeds the configured alarm limit.
Check the patient condition.
XX score = 3
Mediate
The parameter score is 3. Check the patient
condition.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...