13 - 2
2.
RR unit
3.
RR high limit
4.
RR low limit
5.
RR Source: it can be SpO
2
, CO
2
or Manual.
6.
Select to input RR manually: displays only in
Spot Check
mode, or when CO
2
is not
configured.
13.3
Measurement Limitations of SpO2-Sourced RR
The following factors may influence the accuracy of RR measurement sourcing from
SpO
2
:
■
Low perfusion
■
Excessive motion on the measurement site
■
Arhythmia
■
RR signal too weak to be detected in the pleth waveform
13.4
Manually Inputting RR Value
To manually input an RR value, follow this procedure:
1.
Select
in the Resp area. A 60s timer is displayed and starts counting seconds.
2.
Count the total number of Resp of the patient in the past minute.
3.
Input the number and select
.
4.
Select the
Save
quick key to save the data.
13.5
Changing the RR Settings
13.5.1
Enabling SpO
2
Source for RR Measurement
In
Spot Check
mode, you can enable the SpO
2
source. Follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select .
2.
Select
Module
→
Other
.
3.
Switch
SpO2 Calc RR
on.
13.5.2
Enabling RR Alarms
In
Continuous Monitoring
mode, you can enable the RR alarm. Follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select
Alarm
tab.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...