Screws
9. Install the Anesthesia Gas Filter Canister
8. Install the Tray or Weigher (Optional)
10. Connect the AGSS System or Passive Scavenging Pipe (Optional)
a. Place the anesthesia gas filter canister (self-prepared) on the weigher or the tray by
aligning the groove.
b. Use the silicone tube to connect the gas filter canister with the AGSS inlet pipeline
connector.
AGSS inlet pipeline connector
Silicone tube
Anesthesia gas filter
canister
AGSS support rail
Stop pin
AGSS gas supply
hose
ools:
●
M4X12 screws:2
●
Phillips screwdriver
7. Install the Oxygen Generator Tray and Connector (Optional)
a. Fix the oxygen generator tray assembly to the column and trolley by using screws and the
fixing plate respectively.
b. Rotate the oxygen generator connector clockwise into the gas supply inlet and connect the
oxygen generator with a tube.
Tray assembly
a. Insert the tray or weigher (Veta 5 Only) into the screw arbor and tighten the screws.
b. Connect the weigher cable with the communication interface (SP).
c. Use three silicon plugs to attach the cable to the column and fix the cable on the back of
the machine.
Knob
a. Rotate the two stop pins into the AGSS support rail and fix the AGSS support rail into the
side of column with screws.
b. Insert the AGSS assembly into the AGSS support rail and rotate the knob tightly. Use the
AGSS gas supply hose to connect the waste gas outlet with the AGSS assembly inlet.
c. Connect the waste gas outlet with the passive scavenging pipe when using the passive
scavenging method.
Passive scavenging pipe
Veta 3/Veta 5 Installation Guide
Screws
Tools:
●
M4X12 screws:6
●
Phillips screwdriver
Tools:
●
M4X12 screws:4
●
Phillips screwdriver
0 4 6 - 0 2 2 5 1 5 - 0 0 ( 1 . 0 )