Display Area, Indicator, and Controller
Specifications
2 - 12
2.6.3
Alarm Lamp
• The monitor provides 2 alarm priority levels: high level alarms or low level alarms.
• The Alarm lamp is on the front panel of the monitor. It is a group of (red/yellow) LED
indicators.
• During a high alarm, the red LED, of which the duty cycle is 50%, flashes at a frequency
of 1.4 – 2.8 Hz (84 times per minute).
• During a low alarm, the yellow LED illuminates continuously.
• In case a high alarm and a low alarm occur simultaneously, only the red LED flashes.
NOTE:
The audio alarm indication complies with EN60601-1-8.
2.6.4
Keys
There are 14 keys on the front panel of the monitor.
KEYS
DESCRIPTION
On/Standby indicator
Turn the monitor on, off, or switch to standby. In the operating state,
press and hold for less than 1 second to switch the device to the
standby or press and hold for 2s to turn the device off.
Start NIBP
Press to start an NIBP measurement.
Deflate
During an NIBP measurement, stops the measurement. Also, suspends
interval mode until the next time NIBP Start is pressed.
Start recording trend data/real-time waveform. Press again to stop
recording. When recording trend data, press and hold for less than 1
second to record the current data displaying on the LCD, or press and
hold for 2 seconds to record all data of the current patient.
Silence
This button has dual functions:
Press and hold for less than 1 second to pause an audio alarm
(parameters in alarm continue to flash).
Press and hold for 2 seconds to silence an alarm.
Set Alarms
Sets alarm parameters.
Interval
Sets the measurement mode and time interval. Option: STAT mode,
Off, and interval mode (options: 1, 2, 3, 5,10, 15, 20, 30, 60, 120,
and 240 minute intervals). In the Off mode, perform measurement
manually.
Menu
Press to view the menu options.
Display
Press to switch between the normal views, PLETH waveform.
Patient Info
Press to automatically create a patient ID.
Patient Size
Press to change patient size
Up Arrow
Press to perform menu operations on the LCD.
OK
Down Arrow
Summary of Contents for Rosebud Accutorr V