SpO
2
Measurement
Specifications
2 - 8
2.4.2.2
Pulse Rate
Resolution:
1 BPM
Update frequency:
1 Hz
Range and Accuracy:
2.4.2.3
Alarm Setting Range
PATIENT TYPE
PR RANGE
ACCURACY IN
STATIC
CONDITION
1
ACCURACY IN
MOTION
CONDITION
2,3,4
Adult/Pediatric/Neonate
25 – 240 BPM
± 3 digits
± 5 digits
1
The Masimo MS-2013 pulse oximeter with an LNOP- Adt sensor was validated for no motion accuracy
in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70% to
100% SpO
2
against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus
one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
2
The Masimo MS-2013 pulse oximeter with LNOP- Adt sensor has been validated for motion accuracy in
human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing
and tapping motions at 2 to 4 Hz. At an amplitude of 1 to 2 cm and non-repetitive motion between 1 to
5 Hz at an amplitude of 2 to 3cm in induced hypoxia studies in the range of 70% to 100% SpO
2
against
a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard devia-
tion. Plus or minus one standard deviation encompasses 68% of the population.
3
The Masimo MS-2013 pulse oximeter with LNOP-Neo and Neo Pt sensors has been validated for neona-
tal motion accuracy in human blood studies on neonates while moving the neonate’s foot at 2 to 4 Hz
at an amplitude of 1 to 2 cm against a laboratory co-oximeter and ECG monitor. This variation equals
plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the
population.
4
The Masimo MS-2013 pulse oximeter has been validated for low perfusion accuracy in bench top test-
ing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater then
0.02% and transmission of greater than 5% for saturations ranging from 70 to 100%. This variation
equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68%
of the population.
SPO
2
MODULE
SPO
2
PR
MS-2013 SpO
2
module
50 – 100
20 – 250 bpm