Specifications
NIBP Measurement
In single fault conditions, the hardware controls the cuff deflation to prevent the cuff
pressure from exceeding the following ranges:
2.2.4
Cuff Inflation
The Non-Invasive Blood Pressure inflation source brings a volume of 200 ccs to a pressure of
300 mmHg in less than or equal to 10 seconds
2.2.5
Maximum Leakage
The Non-Invasive Blood Pressure Cuff driver allows a pressure drop to be, at most, 1 mmHg
in 10 seconds as measured with a 200 cc volume at differential pressures of 250 mmHg,
150 mmHg, and 50 mmHg.
2.2.6
Deflation Rate
The Non-Invasive Blood Pressure Cuff venting mechanism in adult mode vents a volume of
500 cc, reduced from a pressure of 260 mmHg to a pressure of 15 mmHg in a maximum of
10 seconds.
The Non-Invasive Blood Pressure Cuff venting mechanism in neonatal mode vents a volume of
500 cc, reduced from a pressure of 150 mmHg to a pressure of 5 mmHg in a maximum of
5 seconds.
2.2.7
Update Period
The Non-Invasive Blood Pressure Parameter has an update period of one (1) second.
2.2.8
Initial NIBP Pressure Setup and Range
The initial pressure is adjustable, and should be set to the following default values:
Adult
330 mmHg
Pediatric
220 mmHg
Neonate
165 mmHg
PATIENT TYPE
DEFAULT INITIAL PRESSURE
ADJUST RANGE
Adult
180 mmHg
100 – 280 mmHg
Pediatric
140 mmHg
60 – 180 mmHg
Neonate
100 mmHg
40 – 120 mmHg
Manual pressure changes in increments of 5 mmHg.