Software upgrade
Test and Calibration
6 - 24
6.12
Software upgrade
CAUTION:
Disconnect the monitor from the patient and make sure that
important data is saved before upgrading the monitor.
CAUTION:
Do not shut down or power off the equipment when
upgrading the bootstrap program. Otherwise, it may cause
the equipment to break down.
CAUTION:
Program upgrades should be performed by qualified
service personnel only.
NOTE:
After upgrading the boot program, re-upgrade the system
program and other programs to ensure compatibility.
NOTE:
Make sure the version of the upgrade package is the correct
one. To obtain the latest upgrade package, contact Service.
The following is a list of upgrade programs:
• Bootstrap program
• System program
• Bar code recognition configuration
• Multilingual library
• General configurations (including passwords, company logo)
• System functional configuration
• FPGA program
• Parameter module programs: SpO
2
module (DPM), NIBP module and optional
Temperature module.
1.
In the
MAINTENANCE
dialog (see FIGURE 6-19), check the installed software version
and compare it with the currently released version to determine if an upgrade is needed.
FIGURE 6-19 MAINTENANCE
dialog
2.
Download the upgrade software through a network to a portable PC or desktop PC.
Connect the PC to the monitor using a cable plugged into the network connectors.
3.
Run the upgrade software.
Summary of Contents for Rosebud Accutorr V