Input/Output Communications
Appendix
19 - 28
Passport V Operating Instructions
19.9
Input/Output Communications
ECG Analog Output Specification
Arterial Blood Pressure Analog Output Specification
Defib Sync Pulse
Bandwidth (-3 dB referenced to 10 Hz):
0.5 to100Hz
Maximum Transmission Delay:
≤
25 ms (with notch on)
Sensitivity (referenced to 10 Hz):
1 V/mV of input,
±
10%
Pacer Enhancement:
Amplitude: +5 V±5 %. Width: 2 ±10 % ms.
Bandwidth (-3 dB referenced to 1 Hz):
0~12.5Hz
Maximum Transmission Delay:
<25 ms
Sensitivity:
1 V/100 mmHg, ±4%
(reference frequency 1Hz)
Connector:
BNC AUX connector (Meets the requirements
of EC60601-1 for short-circuit protection and
leakage current.)
Max time delay:
35 ms (R-wave peak to leading edge of pulse)
Amplitude:
High level: 3.5~5V, providing a maximum of
1 mA output current;
Low level: <0.5V, receiving a maximum of 5
mA input current.
Pulse Width:
100 ms
±
10 %
Output Impedance:
50
Ω
Limited Current:
15mA rating
Rising and Falling Time:
<2ms