Prompt Messages
Messages
14 - 16
Passport V Operating Instructions
14.3.8
GAS Prompt Message
Caused by Measurement
Error
Measurement Error
Repeat calibration
procedure. If problem
persists, contact
Technical Support.
Oridion
Caused by no stable gas flow
No stable gas flow
Check the gas
connections or gas
concentration. Repeat
calibration procedure.
If problem persists,
contact Technical
Support.
Oridion
ALARM MESSAGE
REASON
SOLUTION
GM: Agent Uncalibrated
Appears after an unsuccessful
calibration attempt of the agent
sensor. The numeric data for all
agents will appear as XXX, and
the agent waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem
persists, contact Technical
Support.
GM: CO
2
Uncalibrated
Appears after an unsuccessful
calibration attempt of the CO
2
sensor. The numeric data for
CO
2
will appear as XXX, and
the CO
2
waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: N
2
O Uncalibrated
Appears after an unsuccessful
calibration attempt of the N
2
O
sensor. The numeric data for
N
2
O will appear as XXX, and
the N
2
O waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: O
2
Uncalibrated
Appears after an unsuccessful
calibration attempt of the O
2
sensor. The numeric data for O
2
will appear as XXX, and the O
2
waveform will be a
flatline.
Ensure proper gas mixture is
attached tightly and regulator
is on. Repeat calibration
procedure. If problem persists,
contact Mindray
Technical Support.
GM: Agent Zero Error
Appears when the system has
been unable to successfully zero
the anesthetic agent sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
GM: CO
2
Zero Error
Appears when the system has
been unable to successfully zero
the CO
2
sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
GM: N
2
O Zero Error
Appears when the system has
been unable to successfully zero
the N
2
O sensor.
Manually start zeroing the
system again. If problem
persists, contact Mindray
Technical Support.
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...