Passport V Operating Instructions
13 - 3
Alarms
Alarms
13.1.2
Alarm Limits
A separate table of alarm limit settings is maintained for each patient size. When the patient
size is changed, the corresponding selections are automatically used. See table below for
alarm ranges. Default settings appear in bold text
FIGURE 13-3
Alarm Settings Menu - Set Limits
Alarm Delay
None
(default)
1 second
2 seconds
3 seconds
4 seconds
5 seconds
6 seconds
7 seconds
8 seconds
Select to adjust audio alarm delay time.
Apnea Delay
10 seconds
15 seconds
20 seconds
(default)
25 seconds
30 seconds
35 seconds
40 seconds
45 seconds
50 seconds
55 seconds
1 min
Select to adjust apnea delay time.
ST Alarm Delay
30 seconds
(default)
45 seconds
1 min
90 seconds
2 min
3 min
Select to adjust ST alarm delay time.
Alarm Setup Menu
MENU ITEM
SELECTIONS
COMMENTS
*
These selections will not be available if the language chosen in the Installation Menu is French.