Panorama® Central Station interface:
Changing alarm limits at the Panorama®
Central Station or WorkStation will automatically change alarm limits on the
DPM 6 or DPM 7 bedside monitor. Changing alarm limits at the DPM 6 or
DPM 7 bedside monitor will automatically change alarm limits on the
Panorama® Central Station and WorkStation.
Adjust Alarm Responses
To adjust a patient’s alarm responses on the DPM 6 or DPM 7 bedside monitor:
a. Select parameter’s waveform or numeric tile to open parameters Setup
menu
b. Select
Alarm Setup >>
c. Select
•
Level
– to adjust the audio and visual alarm indicators. Level options include
High (Red - flashing), Medium (Yellow - flashing), and Low (Yellow - no flashing)
•
Record
– to activate an automatic recording when the alarm occurs
•
Default
– to restore preset settings
d. Select
Main Menu
(quick key) or
to return to the main display
Or
a. Select
Alarm
Setup from
(quick key) or
Main Menu
(quick
key)
b. Select
Parameters
tab to change parameters or
ST
tab to change ST
c. Scroll and select parameter
d. Select
•
Level
– to adjust the audio and visual alarm indicators. Level options include
High (Red - flashing), Medium (Yellow - flashing), and Low (Yellow - no flashing)
•
Record
– to activate an automatic recording when the alarm occurs
•
Default
– to restore preset settings
e. Select
Main Menu
(quick key) or
to return to the main display
Panorama® Central Station interface:
Changing alarm responses at the
Panorama® Central Station or WorkStation will not change alarm responses on
the DPM 6 or DPM 7 bedside monitor. Changing alarm responses at the DPM 6
or DPM 7 bedside monitor will not change alarm responses on the Panorama®
Central Station and WorkStation.