background image

Quick Reference Guide

Summary of Contents for DPM 6

Page 1: ...Quick Reference Guide...

Page 2: ...t intended as a replacement to the Operating Instructions Prior to operating equipment the user must be familiar with the Operating Instructions Manual contents This document is a guideline only to be...

Page 3: ...able Pacemaker Detection And Set Pacemaker Rate 4 Initiate Relearn 5 Initiate ST Analysis 5 NIBP 5 Initiate a Non Invasive Blood Pressure Measurement NIBP 5 Set a Non Invasive Blood Pressure Measureme...

Page 4: ...t Measurement 11 Initiate Pulmonary Artery Wedge Pressure PAWP 12 Initiate Hemodynamic Calculations 13 Screen Display 14 History 15 Tabular Trends 15 Arrhythmia And Alarm Events 16 Full Disclosure 16...

Page 5: ...ro IBP Change screen Enter the patient setup menu Trigger a manual event Default configurations Start cardiac output procedure Blank Start the real time print Print Setup Have a split screen view of m...

Page 6: ......

Page 7: ...ically enters it at the bedside monitor Entering demographic information at the bedside monitor automatically enters it at the Panorama Central Station and WorkStation Admit is not required for patien...

Page 8: ...nt on the Panorama Central Station will discharge the patient from the WorkStation ViewStation and DPM 6 or DPM 7 bedside monitor optional STANDBY Enter Standby The Standby function allows the clinici...

Page 9: ...M 6 or DPM 7 bedside monitor s integrated recorder a Main Menu quick key b Select Record Setup menu c Select Waveform 1 Waveform 2 or Waveform 3 d Select the waveforms for recording from drop down lis...

Page 10: ...monitor Panorama Central Station WorkStation and ViewStation are independent Enable Pacemaker Detection And Set Pacemaker Rate The DPM 6 and DPM 7 allow the clinician to enable pacemaker detection The...

Page 11: ...Select Main Menu quick key or to return to the main display NIBP Initiate a Non Invasive Blood Pressure Measurement NIBP To initiate a NIBP measurement on a patient on the DPM 6 or DPM 7 bedside monit...

Page 12: ...Or a Select NIBP numeric tile or select NIBP Setup from the Parameters menu in Main Menu quick key to enter the NIBP Setup menu b Select Stop All c Select Main Menu quick key or to return to the main...

Page 13: ...and alarm responses on the DPM 6 or DPM 7 bedside monitor To adjust a patient s alarm limit on the DPM 6 or DPM 7 bedside monitor a Select parameter s waveform or numeric tile to open parameters Setup...

Page 14: ...ate an automatic recording when the alarm occurs Default to restore preset settings d Select Main Menu quick key or to return to the main display Or a Select Alarm Setup from quick key or Main Menu qu...

Page 15: ...isplay Panorama Central Station interface Adjusting arrhythmia alarms at the Panorama Central Station or WorkStation will not change arrhythmia alarms on the DPM 6 or DPM 7 bedside monitor Adjusting a...

Page 16: ...y on front panel of bedside monitor Panorama Central Station interface Selecting Alarm Pause at the DPM 6 or DPM 7 bedside monitor will silence active alarms on the Panorama Central Station WorkStatio...

Page 17: ...ta within a Panorama Central Station or to a different Panorama Central Station may affect the DPM 6 or DPM 7 bedside monitor HEMODYNAMIC MONITORING Initiate A Cardiac Output Measurement Cardiac Outpu...

Page 18: ...the Hemodynamic Calculations menu Note Blood temperature alarms are inactive when a cardiac output measurement is initiated Initiate Pulmonary Artery Wedge Pressure PAWP Pulmonary Artery Wedge Pressur...

Page 19: ...ion Initiate Hemodynamic Calculations Hemodynamic calculations are a set of values that are used to determine the hemodynamic status of the patient To initiate hemodynamic calculations on the DPM 6 or...

Page 20: ...veforms and numeric parameters Configurable on the main display OxyCRG Screen Split screen with OxyCRG HR trend SpO2 trend and RR trend RR waveform on the lower portion of the display with real time w...

Page 21: ...t of the DPM 6 or DPM 7 bedside monitor HISTORY Tabular Trends To display Tabular trends on a patient on the DPM 6 or DPM 7 bedside monitor a Select Review menu quick key or from Main Menu quick key b...

Page 22: ...ect event d Select Details to view the 8 second event e Select Record to print to the integrated recorder Print to print to the independent printer f Select Main Menu quick key or to return to the mai...

Page 23: ...or and Panorama Central Station are independent VOLUMES The alarm volume key volume and QRS volume can be adjusted on the DPM 6 or DPM 7 bedside monitors To adjust the alarm volume on the DPM 6 or DPM...

Page 24: ...urce When monitoring SpO2 there is a variable pitch tone which changes as the patient s saturation level changes The pitch of the tone rises as the saturation level increases and falls as the saturati...

Page 25: ...n to the main display Panorama Central Station Interface Parameter color on the DPM 6 or DPM 7 monitor Panorama Central Station ViewStation and WorkStation are independent DEFAULT PRESET SETTINGS The...

Page 26: ...Mindray DS USA Inc 800 MacArthur Blvd Mahwah NJ 07430 Tel 1 800 288 2121 or 201 995 8000 www na mindray com P N 046 003757 00 Rev A...

Reviews: