PW/CW Mode
1. Press
[PW]/ [CW]
to enter PW/CW mode.
2. Move
[Trackball]
to change the position of sample volume.
3. Change the size and angle of sample volume via Touch screen or its relative knob. Press
[Update]
or
[PW]/ [CW]
to get the
Pulse Wave Doppler.
Image Adjustment
1. Mode displaying areas: switches the active imaging mode.
2. Special modes available under current condition.
3. Parameter adjusting areas: Displays parameters in the current
imaging mode.
4. Knob adjusting areas: Parameters displayed on the bottom of
Touch screen can be adjusted by the knobs under the Touch screen.
Press
[B]
knob to return to B mode.
Parameters in B mode
: Depth, TGC, Gain, iTouch, Focus Position, Steer, Image Quality, Tint Map, U/D (L/R) Flip, Rotation, FOV,
ExFOV, Image Merge, Dynamic Range, Gray Map, Line Density, Focus Number, iClear, TSI, Persistence, iBeam, HScale, Dual Live,
LGC, A. power.
Parameters in Color mode
: Gain, Depth, iTouch, Steer, Scale, Baseline, Invert, B/C Align, Dual Live, Image Quality, Flow State,
Priority, Packet Size, Color Map, Wall Filter, Line Density, Smooth, Persistence, Velocity Tag.
Parameters might be displayed on different pages.
Special Imaging Mode (Optional)
Elastography
1. Select the exam mode (general application for Breast or Thyroid) via Touch screen with any linear probe.
2. Press
[Elasto]
on the control panel to activate the function and press the probe on the target organ.
[Note]: There is a pressure indicator on the bottom of the screen, make sure the pressure you applied is appropriate so that the
pressure bar is indicated in green; otherwise the pressure bar is indicated in gray.
Tissue Doppler Imaging
1. Press
[TDI]
to enter the corresponding TVI mode, parameters will be
displayed on the Touch screen.
2 Press
[M]
button to enter TVM mode; press
[PW]
button to enter TVD
mode; press
[Power]
to enter TEI mode.