2 - 17
Installation Guide
Installation Checkout Procedure
15. Oxygen Sensor Calibration
NOTE:
Oxygen Sensor Calibration can be performed in all
ventilation modes.
NOTE:
See ‘‘Periodic Maintenance Schedule of Service Activities’’
on page 6-2 for when to calibrate the oxygen sensor.
a.
Allow the breathing system to warm up and reach thermal equilibrium
(approximately 30-60 minutes).
b.
Press the
MENU
key and then use the
Navigator
™
Knob
to scroll to the
Calibrate
menu tab (see FIGURE 2-13). Select the
Start Calibration
button.
c.
After the
Start Calibration
button has been selected, the screen shown in
FIGURE 2-14 or FIGURE 2-16will be displayed, instructing the user to remove the
oxygen sensor from the Breathing System and expose it to room air for at least 3
minutes before proceeding. The O
2
sensor voltage is displayed during the
calibration. This is the amplified O
2
cell voltage at the A/D converter for the oxygen
sensor.
NOTE:
The O
2
sensor voltage is not displayed for UI versions 2.24
and lower.
d.
Flush the O
2
sensor with air from the auxiliary output for 5-10 seconds to ensure that
no O
2
bubbles are trapped in the sensor.
NOTE:
Do not shake the O
2
sensor during calibration.
NOTE:
Keep the O
2
sensor in a vertical position, connector side up,
during calibration.
NOTE:
Place the O
2
sensor on top of the heated block during
calibration to minimize the temperature difference from
within the heated block.
FIGURE 2-13 Calibrate
Menu Tab
Summary of Contents for Datascope AS 3000
Page 2: ...AS Datascope Anesthesia Delivery System Service Manual...
Page 9: ...Table of Contents vi 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 13: ...Introduction Notes x 0070 10 0683 AS3000 Service Manual This page intentionally left blank...
Page 179: ...0070 10 0683 Rev K August 8 2014...