17 - 4
BeneVision N1 Patient Monitor Operator’s Manual
3.
Measure or manually enter other required parameters and observations.
4.
If enabled,
select the clinician ID
.
5.
Select
Calculate
to get the total score.
6.
If
Score Confirmation
is enabled, select
Confirm
to save current scoring, or select
Cancel
to give up
current scoring. Refer to section
17.1.4.2 Setting the Scoring Confirmation Switch
for more information.
CAUTION
•
The decision to use Scale 2 of the SpO2 Scale should be made by a competent clinical decision maker
and should be recorded in the patient’s clinical notes.
NOTE
•
Before calculating the score, select Reset to clear the previous score.
•
The keyboard symbol at the right of the parameter value indicates that the value is manually
entered.
•
You can get the score only when all required parameters have been measured or entered.
17.1.4
Changing EWS Settings
17.1.4.1
Changing the Scoring Protocol
The monitor is configured with a default scoring protocol. To change the scoring protocol, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Score
.
17.1.4.2
Setting the Scoring Confirmation Switch
To select if confirmation is required before saving score, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Scoring Confirmation
switch.
◆
Off: the monitor automatically saves the scoring result after the scoring is completed.
◆
On: you need to confirm that whether the scoring result is saved or not after the scoring is completed.
17.1.4.3
Setting the Manual Data Timeout
The manually input parameter data become invalid after a preset time. To set the timeout period for the input
data, follow this procedure:
1.
From the EWS screen select
Setup
.
2.
From the
Manual Data Timeout
area, select a desired parameter and set its timeout period.
NOTE
•
If the data is expired and not updated, the monitor displays the corresponding parameter score in
outline font, and gives a timeout alarm.
17.1.4.4
Setting Auto Scoring
The monitor automatically starts scoring at the preset interval. To enable auto scoring, follow this procedure:
1.
From the EWS page select
Setup
.
2.
Set
Auto Scoring
:
◆
Interval
: the monitor automatically starts scoring at the preset interval.
◆
NIBP
: the monitor automatically starts scoring at the completion of each NIBP measurement.
◆
Alarm
: the monitor automatically starts scoring when an alarm occurs to the parameter for scoring.