Monitoring SpO
2
8-7
8.5.6 Setting the Pulse Rate
In the [
SpO2 Setup
] menu, select [
PR Setup
] to enter the [
PR Setup
] menu. In the [
PR
Setup
] menu, you can:
Select [
PR
] and then select either:
[
On
] to display the PR parameter window, or
[
Off
] to hide the PR parameter window.
Select [
Alm Source
] and then select either:
[
HR
]: if you want the HR to be the alarm source.
[
PR
] if you want the PR to be the alarm source.
[
Auto
]: If the [
Alm Source
] is set to [
Auto
], the monitor will use the heart rate
from the ECG measurement as the alarm source whenever a valid heart rate is
available. If the ECG signal is too weak, the monitor will automatically switch to
PR as the alarm source. As soon as the ECG signal recovers, the monitor will
automatically switch back to the heart rate as the alarm source.
8.5.7 Changing the Speed of the Pleth Wave
In the [
Pleth Waveform
] menu, select [
Sweep
] and and then select the appropriate setting.
The faster the waveform sweeps, the wider the waveform is.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...