4.2.2.2 Infusion Pump with Silicone Specified Infusion Sets
System will inspect whether the set is loaded after completing the self-test:
If infusion set is not loaded, enter the infusion set [
Loading Guide
] interface; If
the infusion set is not required to load, press
to skip the step.
If conventional infusion set is loaded correctly or silicone specified infusion set is
loaded incorrectly, and door is closed, enter the [
Loading Error
] or [
Loading
Guide
] interface.
If silicone specified infusion set is correctly loaded and door is closed, and [
User
Maintenance
]→[
Brand Selection
]→[
On
], enter the [
Set Selection
] interface; if
the switch of [
Brand Selection
] is [
Off
], then skip the [
Set Selection
] interface.
Load infusion set according to the following method:
1. Pull the door latch up to open the
door.
2. Load the set clips to the locating
slot, in the direction indicated by the
arrow.
3. Adjust the slide clamp, insert it into
the operating lever; load the set,
confirm it is firmly loaded into the tube
slot, and adjust or press the set at the
gray zone, to confirm the set is
loaded correctly.