1-2
1.1.1 Warnings
WARNING
Device, cables and accessories must be inspected before use to
guarantee their normal and safe operation.
This equipment can only be connected to the socket with ground
protection. Please adopt a rechargeable battery instead of the socket as
the power supply if the socket is not provided with a ground lead.
To prevent fire or explosion, do not operate the equipment in the
presence of anesthetic, flammable or explosive materials.
Do not open the equipment casing as there is the impending danger of
electric shock. Equipment maintenance and upgrades must be carried
out by maintenance technicians whom are trained and licensed by the
manufacturer. Moreover, the process must be done only after the AC
power supply is disconnected. Maintenance carried out by individuals
non-affiliated to the manufacturer or by non-licensed personnel may
affect the safety, performance and function of the product.
When used with electrosurgery equipment, the safety of animals should
be ensured.
The animal's clinical condition and the working condition of the infusion
pump must be monitored carefully, the alarm volume and alarm levels
need to be set according to the actual needs. Operation and
performance relying solely on the auditory alarm system alone is not
sufficient, and setting the alarm at a low volume may endanger the
animal. If the alarm volume is less than the surroundings volume, which
can further lead to operators identify alarm mistakenly.
The interconnection of other infusion systems or ACCESSORIES to the
ANIMAL LINE may lead to performance degradation and failure to
achieve the expected performance, the working condition of the pump
and animal's clinical condition shall be monitored regularly.
Please carefully install the power line and cables with various
accessories to prevent the animal from choking or suffocation caused
by entanglement of the cables or by electrical disturbance.
The packaging materials must be disposed of in compliance with local
laws and regulations or the hospital policy on waste management.They
must be kept out of the reach of children.
Infusion set knots, filter coagulation and occlusions arising from needle
insertion can cause the pressure inside the infusion set to rise during
infusion. When this occurs, removing the occlusion can cause
excessive liquid to be infused into the animal, so appropriate measures
Summary of Contents for BeneFusion VP3 Vet
Page 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Page 2: ......
Page 73: ...A 11 Unit of Rate ml h...
Page 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Page 91: ......
Page 92: ......