5-3
5.2.2 Removing the Push and Pull Box Assembly (Note: You
should wear an anti-static wristband throughout)
1. As shown in the figure: Pinch the gripper, pull the push and pull box to the middle of
the handle ring, and remove the two M3X6 Phillips pan screws with a Phillips
screwdriver.
2. As shown in the figure: Put the equipment on the side, open the syringe clamp arms,
and use the Phillips screwdriver to remove the four ST2.6x8 Phillips head
self-tapping screws on the inside of the push and pull box.
3. As shown in the figure: Pinch the gripper and push the push and pull box to the
innermost end.