Status and Error Codes
Operation
3 - 48
Accutorr V Operating Instructions
3.17.3
General Alarm Messages of Parameter Modules
3.17.4
NIBP Module Alarm Messages
ALARM MESSAGE
A
B
LEVEL
CAUSE
ACTION
XX INIT ERR N
Yes
No
H
An error occurs to the XX
module during
initialization.
Restart the monitor. If the
problem persists, contact
service personnel for
repair.
NOTE:
Note: N stands for error code.
XX COMM STOP
No
No
H
Failure in communication
between XX module and
the main board.
XX COMM ERROR
No
No
H
Failure in communication
between XX module and
the main board.
XX ALM LMT ERR
No
No
H
The alarm limit of the XX
parameter is changed
inadvertently.
If the problem persists,
contact service personnel
for repair.
XX OUT OF RANGE
No
No
H
The measured XX value
exceeds the measurement
range.
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.
ALARM MESSAGE
A
B
LEVEL
CAUSE
ACTION
NIBP SELFTEST ERR
No
No
L
An error occurs during
NIBP module
initialization.
Select NIBP RESET in the
MAINTAIN menu. If the
problem still exists, contact
service personnel for
repair. See Section 4.8.
NIBP INIT ERR
No
No
L
An error occurs during
NIBP module
initialization.
Restart the monitor. If the
problem still exists, contact
service personnel for
repair.
NIBP COMM ERR
No
No
L
Communication between
NIBP module and the
host fails.
LOOSE CUFF
No
No
L
The NIBP cuff is not
properly connected.
Reconnect the NIBP cuff.
AIR LEAK
No
No
L
The NIBP cuff is not
properly connected or
there is leak in the
airway.
Check the connections or
use a new cuff. If the
problem persists, contact
our service personnel for
repair.
AIR PRESSURE ERROR
No
No
L
Failures occur in the pulse
measurement. The
monitor cannot perform
measurement, analysis,
or calculation.
Check the connections
between the cuff, the
monitor and the patient. or
use a new cuff. It the
problem persists, contact
our service personnel for
repair.
The “A” field indicates whether all alarm indications can be cleared or not, and the “B” field indicates whether all alarm
indications except the alarm message can be cleared or not. The alarm levels are H = high and L = low.