Operating Procedures
4-3
4.3
When the Immersion Method is Used
Water immersion is required when performing an ultrasound exam using transducers,
including 6LE7(s), 6LE7P, 6LB7(s), 6LB7E, 65EL60EA, and 65EL66HA.
The Mindray recommends using syringe: CIVCO 610-224. Transducer 6LB7 is used as
an example to describe the immersion method.
Perform the following operations before using the syringe.
1. Open the syringe pack and remove the syringe.
2. Take out the dustproof cap at the end of the syringe’s extension tube, as shown in
the following figure.
3. Push back the locking nut to extend the connector, as shown in the following figure.
The connector should be connected to the transducer’s injection/drainage port.
Please refer to the following operations when the immersion method is used.
1. Put on the transducer cover.
For the procedures of putting
on the transducer sheath,
refer to subsection above.
Check the transducer sheath
for wrinkles and gaps.