Minco TT269 Installation And Operating Instructions Download Page 3

3

TT269 RTD Temperature Transmitter

Specifications

Input:

  2- or 3-wire 100 ohm platinum RTD's.

Output:

  0 to 1 mA DC over specified range.

Accuracy:

  +/- 0.1% of span when factory calibrated.

Linearity:

  +/- 0.1% of span.

Adjustments:

  3 field selectable temperatures ranges,

Ambient Temperature:

   

Operating:  -40 to 85 °C ( -40 to 185 °F ).

   

Storage:  -55 to 100 °C ( -67 to 212 °F ).

Ambient Temperature Effects:

+/- 0.01% of span/°C.

   

(+/- 0.006% of span/°F).

Supply Voltage:

  10 to 35 volts DC.

Reverse polarity protected.

Voltage effect:

  +/- 0.001% of span per volt.

Lead Wire Compensation (3-wire RTD):

  +/- 0.05% of 

span per ohm, up to 25 ohms in each leg.

Maximum Load Resistance:

  The maximum allowable 

resistance is 1000 

.

Connections:

  Terminal blocks accept wires from AWG

22 to AWG 14.

Physical:

  Epoxy potted for moisture resistance.

Weight:

  2.0 oz. (56.8 grams).

How to Order:

TT269  PD

 

 1  MQ

   . . . Sample part number

                                  

         

     

Temperature Range:

  Choose a code corresponding to the desired temperature range.

                 

RX  = Uncalibrated

              

MQ = 0 to 120 °C

            

TZ  = 0 to 160 °C

           

 

UA  = 0 to 180 °C

                                 

Output:

  1 = 0 to 1 mA DC.

                          RTD Element Code:

                              

PD = 100 

 platinum at 0 °C ( .00385 TCR )

                  Model Number:

  TT269, RTD Temptran.

Dimensions:

  All dimensions are in inches ( millimeters )

When quality and performance are as important as price, call...

 

  PRODUCTS, INC.

7300 Commerce Lane/Minneapolis, Minnesota 55432-3177 U.S.A.
Telephone:(763)571-3121  /   FAX:(763)571-0927