background image

M ilw a u ke e

ISMIL101   11/2001

Follow the same procedure if using
pH10.01 solution.

Adjust the SLOPE trimmer
(pH4/10) on the front panel,
with a small screwdriver, until
the LCD shows the pH value
at the above temperature (see
the pH versus temperature
chart).
E.g. in this case, if the tem-
perature is 15ºC, the meter
display should be adjusted to
read “pH 4.00”.

Calibration is now complete.

pH VERSUS TEMPERATURE CHART:

TEMP

                  pH  VALUES

ºC      ºF

 0        32

    4.01                    7.13                   10.32

 5        41

    4.00                    7.10                   10.24

10      50

    4.00                    7.07                   10.18

15      59

    4.00                    7.04                   10.12

20      68

    4.00                    7.03                   10.06

25      77

    4.01                    7.01                   10.01

30      86

    4.02                    7.00                    9.96

35      95

    4.03                    6.99                    9.92

40     104

    4.04                    6.98                    9.85

45     113

    4.05                    6.98                    9.85

50     122

    4.06                    6.98                    9.82

55     131

    4.07                    6.98                    9.79

60     140

    4.09                    6.98                    9.77

65     149

    4.11                    6.99                    9.76

70     158

    4.12                    6.99                    9.75

M10004B

M10007B

M10010B

MA 9004

MA 9007

MA 9010

BATTERY REPLACEMENT:

When the battery be-
comes weak the meter
will display "V".

When the low battery in-
dicator appears, only a
few hours of battery life
remain. A low battery will
result in unreliable mea-
surements. It is
recommended to replace
the battery immediately.

Battery replacement
must only take place in a
non-hazardous area us-
ing an alkaline 9V battery.

Turn the meter off, slide
off the battery compartment cover at the rear
of the meter and replace the 9V battery with
a new one. Make sure the battery contacts
are tight and secure, seat the battery in its
compartment and insert  the cover again.

OPTIONAL ACCESSORIES:

MA 9004

pH4.01 buffer solution,
220 mL bottle

MA 9007

pH7.01 buffer solution,
220 mL bottle

MA 9010

pH10.01 buffer solution,
220 mL bottle

MA 9015

Electrode storage solu-
tion, 220 mL bottle

MA 9016

General cleaning solu-
tion, 220 mL bottle

M10000B

Rinse solution, 20 ml
sachet (25 pcs.)

MA911B/1

pH electrode with BNC
connector and 1 m cable

MA950

Portable meter wall fixing
kit

SPECIFICATIONS:

SM 101

RANGE

0.00 to 14.00 pH

RESOLUTION

0.01 pH

ACCURACY (@25°C)

±0.02 pH

TEMPERTURE                    Manual setting
COMPENSATION

0 to 50ºC

CALIBRATION

Manual, 2-point

through Offset and Slope trimmers

pH ELECTRODE

MA911B/1
(included)

ENVIRONMENT

0 to 50°C, 95% RH max.

BATTERY TYPE

1 x 9V alkaline (included)

BATTERY LIFE

approx. 70 hours of use

DIMENSIONS

 145 x 80 x 40 mm

WEIGHT

220 g (with battery) meter only

Reviews: