OPERATIONAL GUIDE

 

y

Remove the electrode cap and immerse the 
meter in 

MA9015

 storage solution for 2 hours to 

activate the electrode.

 

y

Turn the meter on by pressing the ON/OFF button. 
All the used segments on the LCD will be visible 
for 1 second or as long as the button is pressed.

 

y

Immerse the electrode in the solution to be 
tested. Stir gently and wait for the reading to 
stabilize, i.e. the instability indicator (hourglass) 
on the LCD turns off.

 

y

Select the pH or ORP (mV) mode by pressing the 
SET/HOLD button.

 

y

The displayed pH value is automatically 
compensated for temperature. The ORP (mV) 
value will be displayed on the main LCD, while 
the temperature will always be displayed on the 
secondary LCD.

 

y

To freeze the display

, while in measurement 

mode, press and hold the SET/HOLD button. 
The “HOLD” message appears on the secondary 
display and the reading will be frozen on the 
LCD. Press any button to return to the normal 
measurement mode.

 

y

To turn the meter off

, press the ON/OFF 

button. The “OFF” message will appear on the 
secondary display. Release the button.

Note:

 

y

Before taking any measurement, make sure the 
meter is calibrated (the CAL tag is on).

 

y

After use always turn the meter OFF, rinse the 
electrode with water to minimize contamination 
and store it with a few drops of storage (

MA9015

or pH7 (

M10007

) solution in the protective cap.

NEVER USE DISTILLED OR DEIONIZED WATER FOR 
STORAGE PURPOSE.

CALIBRATION PROCEDURE

The ORP (mV) range is factory calibrated. Contact 
the nearest Service Center for recalibration, if 
necessary.
For better accuracy, frequent calibration of the tester 
is recommended. Calibration is also necessary after 
electrode replacement, after testing aggressive 
chemicals and where extreme accuracy is required.

 

y

Turn the meter on and select the pH mode with 
the SET/HOLD button.

 

y

Press and hold the ON/OFF/CAL button until 
the “OFF” message on the secondary LCD is 
substituted by “CAL”. Release the button.

 

y

The instrument enters the calibration mode by 
displaying “pH 7.01 USE” (or “pH 6.86 USE” if the 
NIST buffer set was selected).

 

y

For a 

single-point calibration

, immerse the 

electrode in any buffer, i.e. pH 4.01, 7.01 (or 6.86), 
10.01 (or 9.18).

 

y

The meter activates the automatic buffer 
recognition. If no valid buffer is detected, the 
meter keeps the USE indication active for 12 
seconds, and then replaces it with WRNG, 
indicating that the sample being measured is 
not a valid buffer. Otherwise, if a valid buffer is 
detected, then its value is shown on the primary 
display, and REC appears on the secondary LCD.

 

y

If the pH 7.01 (or pH 6.86) was used, press the 
SET button to exit the Calibration mode and the 
“OK 1” message will appear on the display. The 
first calibration point is stored and the meter 
returns to normal measurement mode.

For better accuracy, it is always recommended to 
perform a 2-point calibration.

 

y

For a 

two-point calibration

, immerse the 

electrode in pH 7.01 (or pH 6.86) buffer solution.

 

y

After the first point has been accepted, the 
meter will then ask for the second buffer and 
the message “pH 4.01 USE” will appear.

 

y

Rinse the electrode and immerse it in the 
second solution (pH 4.01, 10.01 or 9.18).

 

y

If a valid buffer value is detected, the REC 
message is displayed and the meter completes 
the calibration procedure. The LCD shows the 
accepted value with the “OK 2” message and the 
instrument returns to the normal measurement 

mode. Otherwise, if no valid buffer is detected, 
the meter displays the WRNG message.

Note:

 When the calibration procedure is completed, 

the CAL tag is turned on.

 

y

To quit the procedure

 and return to the last 

calibration data, after entering the calibration 
mode press the ON/OFF button. The secondary 
LCD displays “ESC” for 1  second and then the 
meter returns to the normal measurement 
mode.

 

y

To reset to the default values

 and clear a 

previous calibration, press the SET/HOLD button 
after entering the calibration mode and before 
the first point is accepted. The secondary LCD 
displays “CLR” for 1 second, the meter resets to 
the default calibration and the CAL tag on the 
LCD turns off.

SETUP

Setup mode allows the selection of temperature (°C 
or °F) and pH buffer set for calibration. To enter the 
Setup mode, press the ON/OFF button until “CAL” on 
the secondary LCD is replaced by “TEMP” and the 
current temperature unit (e.g. TEMP °C). Then:

 

y

for °C/°F selection:

 use the SET/HOLD button; 

then press the ON/OFF button once to enter 
the buffer set selection or twice to return to the 
normal measurement mode.

 

y

to change the calibration buffer set:

 after 

setting the temperature unit, press ON/OFF 
once and select the buffer set (“pH 7.01 BUFF” or 
“pH 6.86 BUFF” for NIST) by using the SET/HOLD 
button. Press ON/OFF to return to the normal 
measurement mode.

ELECTRODE REPLACEMENT

 

y

Remove the protective cap and unscrew the 
plastic ring on the top of the electrode.

 

y

Pull out the 

MI58P

 electrode and replace it with 

a new one.

 

y

Make sure the gaskets are in place before 
screwing back the ring.

BATTERY REPLACEMENT

When the batteries become weak, the battery 
symbol on the LCD will light up to advise that only a 
few hours of working time is remaining.