background image

Main

LCD

Measuring Unit on Main LCD

Instability

Indicator

Battery

Indicator

Temperature Reading 

on Secondary LCD

OPERATIONAL GUIDE

 

y

Remove the electrode and turn the meter 

on by pressing the ON/OFF button. All the 

used segments on the LCD will be visible 

for 1 second or as long as the button is 

pressed.

 

y

Immerse the electrode in the solution to 

be tested.  Stir gently and wait for stable 

reading, i.e. the instability indicator 

(hourglass) on the LCD turns off. 

 

y

The ORP (mV) value will be displayed on 

the main LCD, while the temperature is 

shown on the secondary LCD.

 

y

To freeze the display,

 while in 

measurement mode, press and hold the 

SET/HOLD button. The “HOLD” message 

appears on the secondary display and 

the reading will be frozen on the LCD. 

Press any button to return to normal 

mode.

 

y

To turn the meter off,

 press the ON/OFF 

button. The “OFF” message will appear 

on the secondary display. Release the 

button.

Note: The meter is factory calibrated. 

Contact the nearest Service Center for 

periodic recalibration.

°C/°F SELECTION

 

y

Press and hold the ON/OFF button 

until the secondary LCD shows “TEMP” 

and the current temperature unit (e.g. 

TEMP °C).

 

y

Use SET/HOLD to select the unit (°C or 

°F); then press again ON/OFF to return to 

the normal mode.

ELECTRODE REPLACEMENT

 

y

Remove the protective cap and unscrew 

the plastic ring on the top of the electrode.

 

y

Pull out the MI57P electrode and replace 

it with a new one.

 

y

Make sure the gaskets are in place 

before screwing back the ring.

BATTERY REPLACEMENT

When the batteries become weak, the 

battery symbol on the LCD will light up to 

advise that only a few hours of working 

time is remaining. 
The meter is also provided with BEPS 

(Battery Error Prevention System), which 

avoids any erroneous readings due to low 

battery level by  automatically switching 

the meter off. 
It is recommended to replace the batteries 

immediately. To replace the batteries 

unscrew the battery compartment cap and 

replace all four 1.5V batteries while paying 

attention to their polarity. Make sure the 

gasket is in place before screwing back the 

cap.
Batteries should only be replaced in a 

non-hazardous area using the battery type 

specified in this instruction manual.

ACCESSORIES

MI57P

Replaceable electrode for 

ORP57

MA9015 Electrode storage solution, 

230 ml

MA9016 Electrode cleaning solution, 

230 ml

M10000B Electrode rinse solution, 20 ml 

sachet, 25 pcs.

SPECIFICATIONS

Range

±1000 mV (ORP)

-5.0 to 60.0 °C / 23.0 to 140.0 °F

Resolution

1 mV

0.1 °C / 0.1 °F

Accuracy

(@25 °C)

±2 mV

±0.5 °C / ±1 °F

Typical EMC 

Deviation

±2 mV

±0.3 °C / ±0.6 °F

Calibration

Factory calibrated

Probe

Replaceable MI57P

Environment -5 to 50 °C; 100% RH max.
Battery Type 4 x 1.5V; IEC LR44, A76
Battery Life approx. 300 hours of use
Auto-off

after 8 minutes of non-use

Dimensions 200 x dia 38 mm
Weight

100 g