Measuring unit on main LCD 

Instability indicator 

 

 

Main LCD 

 
Low battery indicator 

Temperature rating 

 

Secondary LCD 

 

OPERATIONAL GUIDE 

• Remove the electrode and turn the meter on by 
pressing the ON/OFF button. All the used 
segments on the LCD will be visible for 1 second 
or as long as the button is pressed. 
• Immerse the electrode in the solution to be 
tested. 
Stir gently and wait for stable reading, i.e. the 
instability indicator (hourglass) on the LCD turns 
off. 
• The ORP (mV) value will be displayed on the 
main LCD, while the temperature is shown on the 
secondary LCD. 
• 

To freeze the display

, while in measurement 

mode,  press and hold the SET/HOLD button. 
The “HOLD” message appears on the secondary 
display and the  reading will be frozen on the 
LCD. 
Press any button to return to normal mode. 
• 

To turn the meter off

, press the ON/OFF button. 

The “OFF” message will appear on the 
secondary display. Release the button. 

Note: The meter is factory calibrated

. Contact 

the 

nearest Service Center for periodic 

recalibration. 

°C/°F SELECTION 

• Press and hold the ON/OFF button until the 
secondary  LCD shows “TEMP” and the current 
temperature unit (e.g. TEMP °C). 
• Use SET/HOLD to select the unit (°C or °F); 
then press again ON/OFF to return to the normal 
mode. 

 

ELECTRODE REPLACEMENT 

• Remove the protective cap and unscrew the 
plastic ring on the top of the electrode. 
• Pull out the 

MI57P 

electrode and replace it with 

a new one. 
• Make sure the gaskets are in place before 
screwing back the ring. 
 

BATTERY REPLACEMENT 

When the batteries become weak, the battery 
symbol  on the LCD will light up to advise that 
only a few hours of working time is remaining. 
The meter is also provided with BEPS (Battery 
Error  Prevention System), which  avoids any 
erroneous readings  due to low battery level by 
automatically switching the meter off. 
It is recommended to replace the batteries 
immediately. 
To replace the batteries unscrew the battery 
compartment  cap and replace all four 1.5V 
batteries while  paying attention to their polarity. 
Make sure the gasket is in place before screwing 
back the cap. 
Batteries should only be replaced in a non-
hazardous  area using the battery type specified 
in this instruction manual. 

 

WARRANTY 

This instrument is warranted against defects in materials  and 
manufacturing for a period of two years from  the date of 
purchase. Probe is warranted for 6 months. 
If during this period the repair or replacement of parts is required, 
where the damage is not due to negligence  or erroneous 
operation by the user, please return the parts to either distributor 
or our office and the repair will be done free of charge. Damages 
due to accidents,  misuse, tampering or lack of prescribed 
maintenance  are not covered. Milwaukee/Martini instruments 
reserves the right to make improvements in design, construction 
and appearance of its products without advance notice.

 