background image

OPERATIONAL GUIDE

• 

Remove the probe cap and turn the meter on 

by pressing the ON/OFF/CAL button. All the 

used segments on the LCD will be visible for 

1 second or as long as the button is pressed.

• 

Immerse the probe in the solution to be tes-

ted and select either pH, EC or TDS mode 

with SET/HOLD button.

• 

Stir gently and wait for the reading to stabilize, 

i.e. the stability indicator on the LCD turns off.

The  pH  and  EC  (or  TDS)  values  are  auto-

matically compensated for temperature and 

will  be  displayed  on  the  main  LCD,  while 

the temperature is shown on the secondary 

LCD.

• To freeze the display

, while in measurement 

mode,  press  and  hold  the  SET/HOLD  but-

ton.  The  “HOLD”  message  appears  on  the 

secondary  display  and    the  reading  will  be 

frozen on the LCD.

• 

Press    any    button  to    return    to    normal 

mode.

• To turn the meter off

, press the ON/OFF/CAL  

button. The  “OFF”  message  will  appear  on 

the secondary display. Release the button.

Notes:

• 

Before  taking  any  measurement,  make  sure 

the meter is calibrated (the CAL tag is on).

• 

After  use  always  turn  the  meter  OFF,  rinse 

the probe with water and store it with the pro-

tective cap.

Measuring unit

on main LCD
Main LCD

Temperature reading 

Secondary LCD

Low battery indicator

Automatic temperature

compensation

Calibration indicator

Stability indicator

SETUP

Setup  mode  allows  the  selection  of  tempera-

ture  (°C  or  °F),  calibration  buffer  set,  in  pH 

mode only,TDS conversion factor (CONV) and 

temperature  coefficient  (BETA),  in  EC  mode 

only. To enter the Setup mode, press the ON/

OFF/CAL button until “CAL” on the secondary 

LCD  is  replaced  by  “TEMP”  and  the  current 

temperature unit (e.g. TEMP °C). Then:

In EC and pH mode:

• 

  for °C/°F selection: 

use the SET/ HOLD button, then press the ON/

OFF/CAL button to go in the settings menu and 

to return to the normal measurement mode.

In EC mode only:

•  

to change the TDS factor value: 

after setting   the temperature unit, press ON/

OFF/CAL button once to show the current val-

ue (e.g. 0.50 CONV). Select the desired value 

by using the SET/HOLD button, then press ON/

OFF/CAL button twice to return to the normal 

measurement mode.

•  

to change the temperature coefficient:

after  setting  the  TDS  factor,  press  ON/OFF/

CAL  button  to  show  the  current  value  of  the 

temperature coefficient ß (e.g. 2.1 BETA). Use 

the SET/ HOLD button to set the desired value, 

then press ON/OFF/CAL button to return to the 

normal measurement mode.

 

In pH mode only:

•  to change the calibration buffer set: 

after  setting  the  temperature  unit,  press  ON/

OFF/CAL button once and select the buffer set 

(“pH 7.01 BUFF” for pH 4.01/7.01/10.01 or “pH 

6.86  BUFF”  for  NIST  set  4.01/6.86/9.18)  by 

using  the  SET/HOLD  button.  Press  ON/OFF/

CAL  button  to  return  to  the  normal  measure-

ment mode.

EC MEASUREMENT AND CALIBRATION 

PROCEDURE

Select  the  EC  or  TDS  mode  with  the  SET/

HOLD button.

Submerge  the  electrode  in  the  solution  to  be 

tested.  The  measurements  should  be  taken 

when the stability indicator 

 on the top left of 

the LCD disappears.

• 

For  better  accuracy,  frequent  calibration  of 

the  tester  is  recommended.  Calibration  is 

also necessary after probe replacement, after 

testing aggressive chemicals and where ex-

treme accuracy is required. From  normal EC 

operation mode, press and hold the ON/OFF/

CAL button until  the  “OFF”  message on the 

secondary LCD is substituted by  “CAL”. Re-

lease the button.

• 

Immerse  the  probe  in  the  proper  calibration 

solution:  M10031  (1413  μS/cm)  for  MW803 

and M10030 (12.88 mS/cm) for MW804.

• 

Once the calibration has been automatically 

performed, the LCD will show “OK” for 1 sec-

ond and the meter will return to normal mea-

surement mode.

• 

Since there is a known relationship between 

EC and TDS readings, it is not necessary to 

calibrate the meter in TDS if it was previously 

calibrated in EC mode.

 

pH MEASUREMENT AND CALIBRATION

Select the pH mode with the SET/HOLD 

button. Submerge the electrode in the solution 

to be tested. The measurements should be ta-

ken when the stability indicator 

 on the top left 

of the LCD disappears.

For better accuracy, frequent calibration of the 

tester  is  recomanded.  Calibration  is  also  ne-

 

cessary after electrode replacement, after tes-

ting aggressive chemicals and where extreme 

accuracy is required.

• 

From normal operation mode, press and hold 

the ON/OFF/CAL button until the OFF message 

on the secondary LCD is substituted by “CAL“. 

Release the button.

• 

The  instrument  enters  the  calibration  mode 

by  displaying  “pH  7.01  USE“  (or  ‘‘pH  6.86 

USE“ if the NIST buffer set was selected).

• 

For  a  single-point  calibration,  immerse  the 

electrode in any buffer, i.e. pH 4.01, 7.01 (or 

6.86), 10.01 (or 9.18).

• 

The  meter  activates  the  automatic  buffer 

recognition. If no valid buffer is detected, the 

meter keeps the USE indication active for 12 

seconds, and then replaces it with WRNG in-

dicating  that  the  sample  being  measured  is 

not a valid buffer. Otherwise, if a valid buffer 

is  detected,  then  its  value  is  shown  on  the 

main  display,  and  REC  appears  on  the  se-

 

condary LCD.

• 

If the pH 7.01 (or 6.86) was used, press the 

SET button to exit the Calibration mode and 

the  “OK  1”  message  will  appear  on  the  dis-

play. The  calibration  point  is  stored  and  the 

meter returns to normal measurement mode.

For better accuracy, it is always recommended 

to perform a 2-point calibration.

• 

For  a  2-point  calibration,  immerse  the  elec-

trode in pH 7.01 (or 6.86) buffer solution.

• 

After  the  first  point  has  been  accepted,  the 

meter will then ask for the second buffer and 

the message “pH 4.01 USE” will appear.

• 

Rinse the electrode and immerse it in the se-

cond solution (pH 4.01,10.01 or 9.18).

 

Notes: 

• 

When  the  calibration  procedure  is  completed, 

the CAL tag is turned on.

• 

To  quit  the  procedure  and  return  to  the  last 

calibration data, after entering the calibration 

mode press the ON/OFF/CAL button. 